Stress‐Induced Degradation Profiling of Evogliptin by Stability‐Indicating Reversed‐Phase High‐Performance Liquid Chromatography and Liquid Chromatography–Mass Spectrometry
Bhairavi Saraf, Preeti Mehta, Rahul Somani
Key points
From the paper's conclusion- These findings support stress association but not definitive structures, as product‐ion MS/MS, accurate‐mass measurements, and authentic degradation‐product standards were not used.
Abstract
A stability‐indicating reversed‐phase high‐performance liquid chromatographic method was developed and validated for quantitative determination of evogliptin in bulk drug substance and a marketed tablet formulation, followed by supporting liquid chromatography‐mass spectrometry investigation of selected stress‐associated components. Separation was achieved on an Eclipse XDB C8 column (250 × 4.6 mm, 5 μm) using acetonitrile and 0.1% formic acid in water (80:20, v/v) at 1.0 mL/min, with photodiode‐array detection at 268 nm and a 10 min run time. The method was linear over 10–60 μg/mL ( R 2 = 0.9993), with repeatability and intermediate precision below 2%. Mean recoveries ranged from 99.657 to 100.555%. The marketed tablet assay was 100.161 ± 0.333%. Forced‐degradation studies showed parent‐drug response losses of 10.90%, 23.61%, and 9.08% under acidic hydrolysis, alkaline hydrolysis, and oxidative stress conditions, respectively. Thermal and photolytic solid‐state exposure produced only 0.40% and 0.08% response loss, respectively, under the conditions investigated. Full‐scan LC–MS showed ions at m/z 186.2, 363.3, and 417.4 associated with DP1, DP2, and DP3, respectively. These findings support stress association but not definitive structures, as product‐ion MS/MS, accurate‐mass measurements, and authentic degradation‐product standards were not used. This study integrates quantitative stability‐indicating chromatographic assessment with conservative mass‐spectrometric investigation of selected stress‐associated components.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Development and Validation of a Stability‐Indicating LC Method for the Determination of Empagliflozin‐Related Substances With Sensitivity Profiling and Robustness Assessment Using Analytical Quality by Design PrinciplesSeparation Science Plus · 2026
- Eco-friendly stability-indicating HPLC method development, validation, and LC-MS-based characterization of degradation products of the polar drug imeglimin.Journal of pharmaceutical sciences · 2026
- Integrated Approach of Risk‐Based Quality by Design and Multicolored Sustainable Analytical Chemistry to Stability‐Indicating Chromatographic Method for Concurrent Analysis of Sitagliptin and GlimepirideSeparation Science Plus · 2025
- Stability indicating green HPLC method for Imeglimin hydrochloride determination in pharmaceutical tablets with comprehensive assessmentScientific Reports · 2026
- Stability-indicating reversed-phase liquid chromatography for fixed-dose combination pharmaceutical products: A systematic review and critical appraisal.Journal of pharmaceutical sciences · 2026
- Development and Validation of a Stability‐Indicating HPLC Method for Degarelix and Characterization of Its Degradation Impurities Using LC–MS, 1 H NMR, and FTIR TechniquesBiomedical Chromatography · 2026