Quantitative and task‐based performance of spectral deep learning reconstruction in low‐dose dual‐layer coronary CT angiography: A phantom study

Lennart R. Koetzier, Stefan J. Fransen, Arnold M. R. Schilham, Marcel J. Greuter, Eran Langzam, Birgitta K. Velthuis, Dominika Suchá, Niels R. van der Werf

Computational

Key points

From the paper's conclusion
  • Deep learning reconstruction (DLR) improves image quality and enables radiation dose reduction in CCTA, but its performance on spectral reconstructions from dual‐layer CT (DLCT) remains unexplored.

Abstract

Background Coronary CT angiography (CCTA) is an established noninvasive test for coronary artery disease, but balancing image quality and radiation dose remains challenging. Dual‐layer spectral CT (DLCT) enables reconstruction of virtual monoenergetic images (VMI) that enhance coronary artery contrast, but are inherently more sensitive to noise, making this trade‐off especially critical. Deep learning reconstruction (DLR) improves image quality and enables radiation dose reduction in CCTA, but its performance on spectral reconstructions from dual‐layer CT (DLCT) remains unexplored. Purpose This phantom study compares spectral DLR at five different noise reduction strengths with hybrid‐iterative reconstruction (hybrid‐IR) at one fixed, clinically used noise reduction strength across conventional and VMI at multiple reduced dose levels. Methods An anthropomorphic thorax phantom with a coronary insert containing two calcifications (100 and 800 mg CaHA/mL) was scanned on a DLCT system at six dose levels (volumetric CT dose index 12.2–1.4 mGy) using a clinical sequential CCTA protocol. Data were reconstructed with one clinical hybrid‐IR protocol, and DLR at five noise reduction strengths (Smoother, Smooth, Standard, Sharp, Sharper) for conventional images and VMI at 7 energy levels (40–70 keV). Image noise, luminal contrast‐to‐noise ratio (CNR lumen ), noise power spectrum means frequency (NPS mean ), task transfer function at 50% (TTF 50 ), and detectability index (d') of coronary plaques were assessed. Statistical significance was inferred when 95% confidence intervals did not overlap. Results At full dose, DLR significantly reduced noise up to 64% in conventional images and 67% in 40 keV VMI, while increasing CNR lumen up to 300% and 230%, respectively, compared with hybrid‐IR. DLR Standard, Sharp, and Sharper significantly increased NPS mean by 7%–13% for conventional images and 40 keV VMI, whereas DLR Smoother shifted noise toward lower frequencies, reducing NPS mean by 22%–32%. All DLR strengths significantly improved spatial resolution of calcified plaque, with TTF 50 increasing 9%–29% relative to hybrid‐IR. VMI at 40 keV with DLR significantly increased d' by 307% for high‐density and 283% for low‐density calcifications compared to conventional hybrid‐IR. At 50% dose reduction, VMI at 40 keV with DLR Smooth improved CNR lumen (+ 46%), and TTF 50 (+ 18%) and d’ (+ 24%) of the high‐density calcification relative to full‐dose hybrid‐IR 40 keV VMI, whereas confidence intervals overlapped for NPS mean and for TTF 50 and d’ of the low‐density calcification. Conclusions On dual‐layer CT, spectral DLR outperforms hybrid‐IR on both conventional and VMI images in noise, CNR, spatial resolution, and plaque detectability. At 50% dose reduction, DLR Smooth on 40 keV VMI exceeded full‐dose hybrid‐IR on 40 keV VMI in CNR, and in spatial resolution and detectability of high‐density calcifications, with overlapping confidence intervals for noise texture and for low‐density calcifications. Patient studies are needed to validate clinical translation.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Quantitative and task‐based performance of spectral deep learning reconstruction in low‐dose dual‐layer coronary CT angiography: A phantom study — Science Skim