Can physics review catch prescription and PTV margin errors in fully automated radiotherapy plans? A hazard study.
Laurence E Court, Raphael Douglas, Parmeswaran Diagaradjane, Vindu W Kathriarachchi, Shiquin S, Ergys Subashi, Yao Zhao, Lifei Zhang,
Department of Radiation Physics, University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
Key points
From the paper's conclusion- Physics plan review was unreliable for detecting incorrect PTV margins in a fully automated planning workflow, even when additional cues were provided. Although based on a limited dataset, these findings indicate that traditional plan review processes alone are insufficient to mitigate this failure mode. Alternative risk-reduction strategies-such as timely workflow-embedded alerts, targeted checklists, automated verification tools, and careful expectation setting-are likely required to improve safety in AI-driven radiotherapy planning systems.
Abstract
PURPOSE: End-to-end AI-based automated contouring and radiotherapy planning systems promise substantial gains in efficiency and consistency but also introduce new safety risks, including user input errors that occur early in the workflow. This study evaluates the ability of standard physics plan review to detect user errors in prescription and planning target volume (PTV) margin entry in a fully automated planning workflow. METHODS: A prototype of the Radiation Planning Assistant (RPA) was used to generate fully automated prostate cancer treatment plans for 20 patients. Five cases intentionally included incorrect prescriptions or non-standard PTV margins. Five clinical medical physicists independently performed physics plan reviews in a commercial treatment planning system, assuming a standard clinical workflow and without access to user guides or service request forms. Reviews were conducted in two phases, with and without margin information embedded in structure names. A third phase evaluated whether adding reference expansion structures improved margin error detection. RESULTS: Incorrect prescriptions were detected in 80% of cases, with no false-positive identification of correct prescriptions. In contrast, detection of incorrect PTV margins was highly inconsistent. Two physicists did not evaluate margins and missed all margin errors. Three physicists identified all intentional margin errors but demonstrated poor specificity, frequently flagging correct cases as incorrect (specificity range: 13%-63%). Including margin information in structure names and adding reference expansion structures did not improve specificity or overall detection performance. CONCLUSIONS: Physics plan review was unreliable for detecting incorrect PTV margins in a fully automated planning workflow, even when additional cues were provided. Although based on a limited dataset, these findings indicate that traditional plan review processes alone are insufficient to mitigate this failure mode. Alternative risk-reduction strategies-such as timely workflow-embedded alerts, targeted checklists, automated verification tools, and careful expectation setting-are likely required to improve safety in AI-driven radiotherapy planning systems.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Clinical impact of introducing automated plan checking in radiation therapy: effects on workflow efficiency, cognitive load and usability.Journal of radiation research · 2026
- Feasibility of quantifying intra-fractional patient positional error for head and neck IMRT using radiomic features of EPID-measured fluence maps: a phantom study.Medical engineering & physics · 2026
- Evidence-based design of a patient-specific quality assurance program through retrospective analysis of pretreatment and in vivo results.Physics and imaging in radiation oncology · 2026
- Automation Bias in AI-Assisted Simulated Radiotherapy Decision-Making Among Undergraduate Radiological Sciences Students: A Controlled Experimental StudyInternational Medical Education · 2026
- An EPID-based virtual audit framework for multi-institutional radiotherapy audits using direct fluence comparison.Physics and imaging in radiation oncology · 2026
- Multi-agent AI system for radiology report structuring and quality assurance with independent radiologist evaluationFrontiers in Radiology · 2026