Impact of MRI slice thickness and fiducial marker orientation on geometric accuracy in multi-echo GRE imaging for prostate radiotherapy planning.
Aika Hayashi, Yasukata Takahashi, Naoko Sanuki, Yasufumi Yamashita, Yoshitaka Suzuki, Kanae Matsuura, Hiroe Muto
Graduate School of Medical Science, Suzuka University of Medical Science, Suzuka, Mie, Japan. · Department of Central Radiology, Yokkaichi Municipal Hospital, Yokkaichi, Mie, Japan.
Key points
From the paper's conclusion- MRI slice thickness and marker orientation significantly impact GFM visualization accuracy. The results suggest that a 1.0 mm slice thickness offers a technical advantage for maintaining sub-millimeter localization accuracy across various marker orientations and highly desirable for high-precision radiotherapy planning.
Abstract
BACKGROUND: Precise target definition in prostate radiotherapy requires accurate computed tomography (CT)-magnetic resonance imaging (MRI) image registration using gold fiducial markers (GFMs). While multi-echo gradient recalled echo (ME GRE) sequences facilitate GFM identification, the quantitative impact of acquisition parameters on physical measurement accuracy remains insufficiently understood. PURPOSE: This study evaluates how MRI slice thickness and marker orientation angle jointly affect GFM visualization and geometric accuracy to characterize the resulting localization uncertainties. METHODS: A kiwifruit-based phantom (T1: 1603.3 ms; T2: 72.3 ms) that mimics the relaxation times of the human prostate was imaged with 3.0T MRI. Imaging was performed using a 3D ME GRE sequence with three slice thicknesses (1.0, 2.0, and 3.0 mm) and 12 orientation angles ranging from horizontal to 90°. Five multidisciplinary observers independently identified marker coordinates while blinded to the acquisition parameters. Subjective confidence scores were also recorded. RESULTS: Two-way analysis of variance (ANOVA) revealed significant main effects for slice thickness (F(2, 144) = 32.03, p < 0.001) and orientation angle (F(11, 144) = 4.06, p < 0.001), with significant interaction between them (F (22, 144) = 2.72, p < 0.001). Mean measurement errors with 95% confidence intervals (CI) were 0.96 mm (95% CI: 0.70-1.22) for 1.0 mm, 1.49 mm (95% CI: 1.23-1.75) for 2.0 mm, and 2.00 mm (95% CI: 1.74-2.26) for 3.0 mm slices. The peak error reached 3.10 mm under the combination of a 3.0 mm slice and 15° orientation. Subjective confidence scores showed a significant negative correlation with physical measurement errors (r = -0.296, p < 0.001). CONCLUSIONS: MRI slice thickness and marker orientation significantly impact GFM visualization accuracy. The results suggest that a 1.0 mm slice thickness offers a technical advantage for maintaining sub-millimeter localization accuracy across various marker orientations and highly desirable for high-precision radiotherapy planning.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- The effect of magnetic resonance image slice thickness and voxel geometry on digital dental implant planning: an in vitro study.Oral and maxillofacial surgery · 2026
- Circumferential Meniscal Extrusion Mapping Using Radial Multiplanar Reconstructions.Journal of orthopaedic research : official publication of the Orthopaedic Research Society · 2026
- Zero echo time versus T1-weighted gradient-recalled echo MRI of breast biopsy markers: a phantom study comparing susceptibility artifact volume and geometric dimensions.Translational breast cancer research : a journal focusing on translational research in breast cancer · 2026
- Treatment-position magnetic resonance imaging reduces bony-landmark co-registration error in head and neck radiotherapy.Physics and imaging in radiation oncology · 2026
- Diagnostic Performance of Magnetic Resonance Elastography Versus Vibration‐Controlled Transient Elastography for Liver Fibrosis Staging in Steatotic Liver DiseaseNMR in Biomedicine · 2026
- Directional dependence of MRI geometric distortion under different receiver bandwidths.Radiological physics and technology · 2026