Treatment-position magnetic resonance imaging reduces bony-landmark co-registration error in head and neck radiotherapy.
Tomohiro Itonaga, Ryuji Mikami, Mitsuru Okubo, Masumi Kawaguchi, Tatsuhiko Zama, Masahiko Kurooka, Yoichi Araki, Shinji Sugahara,
Department of Radiology, Tokyo Medical University Hospital, 6-7-1 Nishi-shinjuku, Shinjuku, Tokyo 160-0023, Japan. · Department of Radiology, Tokyo Medical University Hachioji Medical Center, 1163 Tate-machi, Hachioji, Tokyo 193-0998, Japan.
Key points
From the paper's conclusion- Treatment-position magnetic resonance imaging reduced bony-landmark error in co-registration with planning computed tomography after rigid registration, and the benefit persisted after deformable refinement. Because delineation and dose-volume parameters were not assessed, these findings should not be interpreted as demonstrating improved contouring accuracy.
Abstract
BACKGROUND AND PURPOSE: Routine diagnostic magnetic resonance imaging is acquired without immobilisation, creating co-registration uncertainty with planning computed tomography. We quantified whether treatment-position imaging reduced bony-landmark co-registration error, and whether the benefit persisted after resampling and deformable refinement. MATERIALS AND METHODS: Fifty patients with paired treatment-position and diagnostic magnetic resonance imaging were retrospectively analysed. Each magnetic resonance imaging dataset was separately registered to planning computed tomography. Mean Euclidean distance between corresponding bony landmarks on the computed tomography and magnetic resonance images was compared after rigid registration and after deformable refinement, using Holm-adjusted landmark-wise comparisons. Through-plane resampling was assessed in all 50 patients, and interobserver reproducibility in a 15-patient subset. RESULTS: Treatment-position imaging reduced mean Euclidean distance from 7.5 to 4.1 mm after rigid registration (difference 3.4 mm; 95% confidence interval 2.7-4.1; p = 2.0 × 10-13). The benefit persisted after resampling (4.3 mm) and deformable refinement (3.4 versus 6.0 mm; difference 2.6 mm; 95% confidence interval 2.1-3.1; p = 7.1 × 10-14), although refinement reduced the advantage by 0.8 mm (interaction p = 0.023). Intraclass correlation coefficients were 0.81 for diagnostic imaging and 0.48 for treatment-position imaging (95% confidence interval - 0.05-0.79), with small mean bias and lower treatment-position error in all 15 patients. CONCLUSIONS: Treatment-position magnetic resonance imaging reduced bony-landmark error in co-registration with planning computed tomography after rigid registration, and the benefit persisted after deformable refinement. Because delineation and dose-volume parameters were not assessed, these findings should not be interpreted as demonstrating improved contouring accuracy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparison of subject-to-template registration schemes using CT and MR radiotherapy images with brain lesions.Frontiers in neuroscience · 2026
- Motion-Robust Dental MRI for Imaging of Paediatric Dental Trauma.Dento maxillo facial radiology · 2026
- Vertebral Body Kinematics Measured From T1-Weighted Magnetic Resonance Imaging With Optimized Rigid RegistrationJournal of Biomechanical Engineering · 2026
- Manual CBCT registration by Japanese radiological technologists: agreement with a clinical reference position and exploratory associations with professional experience.Radiological physics and technology · 2026
- Improved correction of anatomical distortion in diffusion-weighted prostate MRImedRxiv · 2026
- P06.16.B Consensus-based white matter tract reconstruction from diffusion MRI acquired in radiotherapy treatment position in primary and secondary brain tumors: interim analysis of a prospective cohortNeuro-Oncology · 2026