Sensor-Free Shockable Rhythm Classification During CPR Using the Existing ECG Channel of Public-Access Automated External Defibrillators
Jaechan Lim, David Hicks, Matt Valentine, Ki H. Chon
Key points
From the paper's conclusion- Thus, rhythm analysis during ongoing CPR can be implemented with the ECG hardware already present in public-access AEDs, without auxiliary sensing hardware, subject to validation on the target device.
Abstract
Automated external defibrillators (AEDs) are public-access devices intended for use by lay rescuers, yet they must pause chest compressions to analyze the rhythm because compressions corrupt the electrocardiogram (ECG). Removing this pause requires a method that uses the hardware already present in the device, fits within its processor budget, and meets the rhythm-recognition performance required for defibrillator approval. Existing approaches rely either on an auxiliary sensor, which entails a hardware redesign, or on an additional network that reconstructs the ECG before classification. This work classifies the compression-contaminated ECG directly, using only the existing ECG channel of the AED. A cascade of three compact convolutional networks trained with different class ratios verifies each decision with a second network biased toward the proposed outcome; disagreement yields an indeterminate outcome. Evaluated on an independent manufacturer-provided cohort of 464 subjects over 15 train-validation partitions, the cascade satisfies both adjudicable American Heart Association criteria under all six noise conditions at an eight-second window and a signal-to-artifact ratio of 0 dB, with sinus-rhythm specificity of 99.20-99.43% and other non-shockable specificity of 95.48-96.20%, leaving only 1.07-1.48% of segments indeterminate. The criteria are not met at -3 dB. Single networks matched in parameter count and inference cost do not reproduce this performance. Each decision evaluates two of the three networks and requires 6.68 ms on a single thread of a general-purpose processor, and eight-bit quantization reduces the model to 1.45 MiB without changing any compliance verdict. Thus, rhythm analysis during ongoing CPR can be implemented with the ECG hardware already present in public-access AEDs, without auxiliary sensing hardware, subject to validation on the target device.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Computer Vision‐Guided Automated External Defibrillator System Improves Cardiopulmonary Resuscitation Accuracy: A Randomized Crossover TrialHealth Science Reports · 2026
- Noninvasive Cardiac Arrest Recognition Using Wearable Sensors in a Porcine Model of Ventricular FibrillationbioRxiv · 2026
- A Fast and Generalizable Deep Neural Network for the Detection of Atrial Fibrillation.Bioengineering (Basel, Switzerland) · 2026
- Diagnostic Performance of a Deep Neural Network Algorithm for Automated Detection of Arrhythmias from Consumer ECG DevicesBIO Web of Conferences · 2026
- RhythmX™: An Interpretable Self-Supervised Contrastive Learning Framework for Heartbeat ClassificationTechnologies · 2026
- Deep learning-based electrocardiography for sudden cardiac death risk prediction: a systematic review and meta-analysis of validation studiesEuropean Heart Journal · 2026