Neural implants and human safety: single-fault detection for DC-coupled recording front ends
Dimitris Antoniadis, Timothy Constandinou
Key points
From the paper's conclusion- Designed in 65 nm, the mechanism resolves a worst-case fault of 6.4 nA across all corners within 0.81 ms - compliant with the ISO~14708-3 limit for an 8533 um2 electrode - opening a broader discussion of safety in DC-coupled recording.
Abstract
DC-coupled analogue front ends (AFEs) for neural implants provide a low-area solution. However, removing the coupling capacitor eliminates the intrinsic barrier that protects cortical tissue: a single-fault event, such as gate-oxide breakdown of a low-noise amplifier (LNA) input transistor, can open a direct DC path from the supply rail into the brain. On the stimulation side this hazard is well understood, and single-fault tolerance is enforced by a series DC-blocking capacitor; on the recording side, DC-coupled front ends discard the equivalent safeguard, yet their protection has gone almost unexamined. This paper presents a single-fault detection mechanism that monitors the LNA for the DC imbalance produced by such a failure and disables the amplifier before the resulting fault current can irreversibly damage tissue. The imbalance is encoded in the duty cycle of a current-starved relaxation oscillator and read out as a time-to-digital measurement. Designed in 65 nm, the mechanism resolves a worst-case fault of 6.4 nA across all corners within 0.81 ms - compliant with the ISO~14708-3 limit for an 8533 um2 electrode - opening a broader discussion of safety in DC-coupled recording.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A subnanolitre tetherless optoelectronic microsystem for chronic neural recording in awake miceNature Electronics · 2025
- A Dual-Mode Neural Amplifier Array for Biopotential and FSCV-Based Neurochemical Measurements.Biosensors · 2026
- A Simultaneous 16-Channel Adiabatic Charge-Balanced Stimulator with 88% Peak Efficiency and 4.1% Variation Across Loads.IEEE transactions on biomedical circuits and systems · 2026
- An Event-Driven Cap-Less Near-Load Digital LDO for Wirelessly Powered Implantable Biomedical SoCs with DICE-FF-Based SAR State-Storage MitigationElectronics · 2026
- High-frequency, low-energy organic event-based sensors for closed-loop neurostimulationNature Sensors · 2026
- NORA: Neuro-headstage for Online Recording and Actuation in Freely Behaving Animal ModelsarXiv · 2026