IEEE transactions on biomedical circuits and systems

A Simultaneous 16-Channel Adiabatic Charge-Balanced Stimulator with 88% Peak Efficiency and 4.1% Variation Across Loads.

Eojin Kim, Geunchang Seong, Hyungwoo Kim, Michal A Gorywoda, Su Yeon Jeon, Seonae Jang, Wanyeong Jung, Young Sang Cho,

Key points

From the paper's conclusion
  • The proposed stimulator achieves a peak energy efficiency of 88% with <4.1% variation over 500 Ω-10 kΩ tissue resistance and 60-250 nC stimulation charge.

Abstract

This paper presents a simultaneous 16-channel adiabatic charge-tracking stimulator (SACT) for implantable neural stimulation systems. Time-multiplexed stimulation enables simultaneous multi-channel operation using a single current-generation block, reducing area scaling with channel count while suppressing charge errors caused by inter-channel crosstalk currents. An on-chip capacitor tracks the delivered stimulation charge for accurate per-channel charge balance, while an inductor adiabatically recovers energy during each stimulation cycle. A low-power predictive phase-timing detector accurately determines phase-completion timing, while an integrated faulty channel detector disables channels under abnormal electrode conditions. The proposed stimulator achieves a peak energy efficiency of 88% with <4.1% variation over 500 Ω-10 kΩ tissue resistance and 60-250 nC stimulation charge. Charge mismatch remains <0.98% during simultaneous two-channel stimulation. The design achieves the smallest projected total area among the compared stimulators for 16 or more channels. In vivo validation demonstrated simultaneous multi-channel stimulation.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

A Simultaneous 16-Channel Adiabatic Charge-Balanced Stimulator with 88% Peak Efficiency and 4.1% Variation Across Loads — Science Skim