Quantitative Imaging of Scandium Isotopes with a Long Axial Field-of-View PET/CT

Lorenzo Mercolli, Mohammadreza Teimoorisichani, Alenka E. Gaschen, Pascal V. Grundler, Anzhelika N. Moiseeva, Saverio Braccini, Nicholas P. van der Meulen

Key points

From the paper's conclusion
  • We evaluated the quantitative imaging performance of $^{44}$Sc and a $^{43/44}$Sc mixture on a long axial field-of-view (LAFOV) PET/CT using a NEMA IEC body phantom.

Abstract

Objective. $^{43}$Sc and $^{44}$Sc are positron-emitting isotopes of scandium that have attracted growing interest for theranostic applications in conjunction with the therapeutic isotope $^{47}$Sc. Both isotopes are non-pure positron emitters, and the associated prompt-gamma emission together with the high positron energy introduce complexities in achieving precise quantitative imaging. Approach. We evaluated the quantitative imaging performance of $^{44}$Sc and a $^{43/44}$Sc mixture on a long axial field-of-view (LAFOV) PET/CT using a NEMA IEC body phantom. We assessed contrast recovery coefficients, image noise and bias for different numbers of iterations and for absolute versus relative scatter correction, and implemented an isotope-specific point-spread-function (PSF) correction based on FLUKA-simulated positron ranges. Main results. The quantitative imaging performance of both isotopes is comparable to $^{18}$F, and the vendor's prompt-gamma correction effectively suppresses the additional random coincidences and lung error caused by prompt-gamma emission. The isotope-specific PSF correction further improves contrast recovery. While the bias obtained for $^{44}$Sc meets the EARL acceptance criterion for non-standard radionuclides, the bias for the $^{43/44}$Sc mixture exceeds it, which we attribute to the mixture-specific cross-calibration of the dose calibrators rather than to the PET imaging. Significance. Our results support the use of LAFOV PET/CT for quantitative imaging of scandium isotopes and provide practical reconstruction recommendations for clinical protocols.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers