Quantitative Imaging of Scandium Isotopes with a Long Axial Field-of-View PET/CT
Lorenzo Mercolli, Mohammadreza Teimoorisichani, Alenka E. Gaschen, Pascal V. Grundler, Anzhelika N. Moiseeva, Saverio Braccini, Nicholas P. van der Meulen
Key points
From the paper's conclusion- We evaluated the quantitative imaging performance of $^{44}$Sc and a $^{43/44}$Sc mixture on a long axial field-of-view (LAFOV) PET/CT using a NEMA IEC body phantom.
Abstract
Objective. $^{43}$Sc and $^{44}$Sc are positron-emitting isotopes of scandium that have attracted growing interest for theranostic applications in conjunction with the therapeutic isotope $^{47}$Sc. Both isotopes are non-pure positron emitters, and the associated prompt-gamma emission together with the high positron energy introduce complexities in achieving precise quantitative imaging. Approach. We evaluated the quantitative imaging performance of $^{44}$Sc and a $^{43/44}$Sc mixture on a long axial field-of-view (LAFOV) PET/CT using a NEMA IEC body phantom. We assessed contrast recovery coefficients, image noise and bias for different numbers of iterations and for absolute versus relative scatter correction, and implemented an isotope-specific point-spread-function (PSF) correction based on FLUKA-simulated positron ranges. Main results. The quantitative imaging performance of both isotopes is comparable to $^{18}$F, and the vendor's prompt-gamma correction effectively suppresses the additional random coincidences and lung error caused by prompt-gamma emission. The isotope-specific PSF correction further improves contrast recovery. While the bias obtained for $^{44}$Sc meets the EARL acceptance criterion for non-standard radionuclides, the bias for the $^{43/44}$Sc mixture exceeds it, which we attribute to the mixture-specific cross-calibration of the dose calibrators rather than to the PET imaging. Significance. Our results support the use of LAFOV PET/CT for quantitative imaging of scandium isotopes and provide practical reconstruction recommendations for clinical protocols.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- First positronium lifetime imaging with scandium-44 on a long axial field-of-view PET/CTFrontiers in Nuclear Medicine · 2025
- Quantitative Comparison of SPECT and PET Performance for Clinical Theranostic ApplicationsJournal of Nuclear Medicine · 2025
- Unlocking the potential of scandium-44: reproducible production and radiolabelling methods towards clinical use in nuclear medicineEJNMMI Radiopharmacy and Chemistry · 2026
- Robustness of Long Axial Field-of-View PET to Defective Detector Blocks: Impact on Quantitative AccuracymedRxiv · 2026
- First Positronium Imaging Using⁴⁴ Sc With the J-PET Scanner: A Case Study on the NEMA-Image Quality PhantomIEEE Transactions on Radiation and Plasma Medical Sciences · 2025
- A Source-Traced PET/CT Scanner Performance Database (2004-2024): The Sphere-Size Dependence of Time-of-Flight Contrast Recovery.Diagnostics (Basel, Switzerland) · 2026