Lorenzo Mercolli, William M. Steinberger, Pascal V. Grundler, Anzhelika Moiseeva, S. Braccini, Maurizio Conti, Paweł Moskal, N. Rathod, Axel Rominger, Hasan Sari, Roger Schibli, Robert Seifert, Kuangyu Shi, Ewa Stępień, Nicholas P. van der Meulen
Purpose The physical properties of 44 Sc, combined with its imminent clinical application, position it as a prime candidate for in vivo positronium lifetime imaging. In this study, we investigate the count statistics for ortho-positronium (oPs) measurements with 44 Sc on a commercial long-axial field-of-view (LAFOV) PET/CT. Method A NEMA image quality phantom was filled with 41.7 MBq of 44 Sc dissolved in water and scanned on a LAFOV PET/CT. Three-photon events were identified using a prototype feature of the scanner and dedicated software. The lifetime of oPs was determined in the phantom spheres and in 4 × 4 × 4 mm 3 voxels. Results All measured oPs lifetimes are compatible, within the uncertainties, with the literature values for water. The oPs lifetime is 2.65 ± 0.50 , 1.39 ± 0.20 and 1.76 ± 0.18 ns in the three smallest spheres of the phantom and 1.79 ± 0.57 ns for a single voxel in the central region of the largest sphere. The relative standard deviation in the background regions of the time difference distributions, i.e., for time differences smaller than − 2.7 ns, is above 20%—even for voxels inside the phantom spheres. Conclusions Despite the favorable physical properties of 44 Sc, the count statistics of three-photon events remains a challenge. The high prompt-photon energy causes a significant amount of random three-photon coincidences with the given methodology and, therefore, increases the statistical uncertainties on the measured oPs lifetime.