Characterization of integrating-sphere systematic effects for photon calibration of the KAGRA gravitational wave detector
Shingo Hido, Misato Onishi, Shinji Miyoki, Takahiro Sawada, Takahiro Yamamoto, Takayuki Tomaru, Yoshiki Moriwaki, Dan Chen
Key points
From the paper's conclusion- These results indicate that variations in the incident-beam conditions of integrating-sphere power sensors account for a substantial part of the previously unexplained scatter and establish a more physically motivated framework for evaluating Pcal laser-power calibration uncertainty.
Abstract
Photon calibration (Pcal) systems are essential for accurately reconstructing gravitational-wave strain in laser interferometric detectors. In the KAGRA Pcal system, the laser power applied to the end test mass is measured using integrating-sphere power sensors. During the first part of the fourth observing run (O4a), repeated calibration measurements showed scatter larger than expected from the uncertainties assigned to individual measurements. Although this excess scatter was empirically reflected in the previously reported calibration uncertainties, its physical origin was not identified. In this study, we experimentally characterize the response of the integrating spheres used in the KAGRA calibration chain with respect to three incident-beam conditions: incidence angle, polarization, and beam position. Based on these measurements and the reproducibility of the integrating-sphere placement in each calibration configuration, we evaluate the corresponding systematic uncertainties and explicitly incorporate them into the uncertainty model for the O4a Pcal-X calibration measurements. The inclusion of these previously unmodeled effects substantially improves the consistency between the estimated uncertainties and the observed scatter, particularly for calibration parameters measured at the KAGRA end station, where the reduced chi-square values decrease from values as large as $10^3$ to values of order unity to several. These results indicate that variations in the incident-beam conditions of integrating-sphere power sensors account for a substantial part of the previously unexplained scatter and establish a more physically motivated framework for evaluating Pcal laser-power calibration uncertainty. They also provide a basis for future reduction of the calibration uncertainty through improved control of the incident-beam conditions.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Improved accuracy of the reference photon calibration standards in the network of gravitational-wave observatories during the O4 observing runarXiv · 2026
- Estimation of systematic error from bulk deformation of end test mass induced by photon calibrator for LIGO post-O5 gravitational wave projectsClassical and Quantum Gravity · 2026
- Accurate calibration method of the Back-linked Fabry Perot interferometer using null channel for space-based gravitational wave antennaarXiv · 2026
- Cyclic-Quadrature Intracavity Signal Amplification for Gravitational-Wave DetectorsClassical and Quantum Gravity · 2026
- Continuously tunable attenuation stage for weak-light radiance calibration using an integrating sphere and a dual-aperture tube.Applied optics · 2026
- Revealing and Mitigating Transmittance and Reflectance Errors in Integrating-Sphere Characterization of High-Albedo Scattering MediaarXiv · 2026