科研速览 · Science Skim继续刷下去 · Keep skimming →
◆ Chemical Science International Journal2026-03-11· Chromatography

Validated Micelle-enhanced Spectrofluorimetric Determination of Serotonin in Human Urine

Abdourahmane Khonté, Khémesse Kital, Jean Pierre Bakhoum, Latyr Dione, Oladele Oyelakin, Diène Diégane Thiaré, Lamine Cisse, Atanasse Coly, François Delattre, Alphonse Tine

原始摘要(英文原文)· Original abstract
A micelle-enhanced spectrofluorimetric method was developed and validated for the determination of serotonin (5-hydroxytryptamine, 5-HT) in aqueous solutions and human urine. Fluorescence responses of 5-HT were evaluated in various surfactant systems, with Brij 700 and sodium dodecyl sulfate (SDS) providing significant signal enhancement. Under optimized conditions (λex 275 nm; λem 340 nm), the method exhibited linearity over 0.1–300 ng mL⁻¹ (R² ≥ 0.996), with limits of detection of 1.7 ng mL⁻¹ in Brij 700 and 2.4 ng mL⁻¹ in SDS. Validation according to ICH Q2(R1) guidelines confirmed good precision (intra-day RSD 1.3–3.6 %, inter-day RSD 3.7–4.3 %), accuracy, selectivity, and reproducibility. Recoveries from fortified urine samples ranged from 89–101 %, demonstrating minimal matrix effects. Compared to conventional chromatographic techniques (HPLC/GC), this micelle-enhanced method is faster, simpler, and cost-effective, making it particularly suitable for routine clinical monitoring of serotonin in human urine.