Intraductal Radiofrequency Ablation Followed by the Stent-in-Stent Technique for the Removal of an Embedded, Intraductally Migrated Fully Covered Biliary Metal Stent: A Case Report.
Felipe S Santos, André O Ardengh, Vitor H Lopes, Renato Baracat, Fauze Maluf-Filho
Gastroenterology and Endoscopy, Hospital das Clínicas da Faculdade de Medicina da Universidade de São Paulo (HC-FMUSP), São Paulo, BRA. · Gastroenterology, Cancer Institute of the State of São Paulo, Hospital das Clínicas da Faculdade de Medicina da Universidade de São Paulo (HC-FMUSP), São Paulo, BRA.
Key points
From the paper's conclusion- We report a 38-year-old woman with a Strasberg E2 bile duct injury after cholecystectomy and a benign segmental stricture of the common hepatic duct, treated with dilation, plastic biliary stents followed by a 10 mm × 6 cm FCSEMS.
Abstract
Fully covered self-expandable metal stents (FCSEMSs) are widely used in benign biliary strictures because they combine sustained radial force with the possibility of later endoscopic removal. When the stent stays in place for too long, tissue hyperplasia and intraductal migration can leave the stent firmly embedded in the duct wall. We report a 38-year-old woman with a Strasberg E2 bile duct injury after cholecystectomy and a benign segmental stricture of the common hepatic duct, treated with dilation, plastic biliary stents followed by a 10 mm × 6 cm FCSEMS. She was subsequently lost to follow-up for two and a half years and returned with mild abdominal pain and cholestasis. By then, the stent had migrated intraductally, with its distal end approximately 2 cm above the major papilla, and was impacted by granulation tissue. Conventional retrieval failed, as did a stent-in-stent attempt with a 10 mm × 8 cm FCSEMS. Cholangioscopy showed granulation tissue within the lumen and at the extremities of the impacted stent. Intraductal bipolar radiofrequency ablation (RFA) was applied at 9 W for 60 seconds along the entire length of the stent using a Habib EndoHPB catheter (Boston Scientific, Marlborough, MA, USA). Hyperplastic tissue was visibly reduced, but the stent could still not be extracted in the same session, so the stent-in-stent technique was repeated with a 10 mm × 8 cm FCSEMS and an 8.5 Fr × 12 cm plastic biliary stent. Finally, 49 days later, the embedded stent had mobilized toward the major papilla and was retrieved with foreign-body forceps. A small, low-output fistula of the mid common bile duct was identified on occlusion cholangiography. It was successfully managed with a further FCSEMS and a plastic stent. Sequential intraductal RFA followed by delayed stent-in-stent placement may serve as a rescue strategy when an embedded biliary FCSEMS is not removable by conventional means or by the stent-in-stent technique alone.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Through-the-scope fully covered self-expandable metal stent with biflanged ends for ampullary access in severe duodenal obstruction.VideoGIE : an official video journal of the American Society for Gastrointestinal Endoscopy · 2026
- Comparative Effectiveness of Two Novel Anti-Migration Strategies in Fully Covered Self-Expandable Metal Stents for Biliary Strictures: A Multicenter Propensity-Matched Study.Gastrointestinal endoscopy · 2026
- Transmural side holes in covered metal stents do not reduce bile duct occlusion or cholecystitis: a single center studymedRxiv · 2026
- Successful endoscopic retrieval of a plastic gallbladder stent that had completely migrated into the gallbladder.Clinical journal of gastroenterology · 2026
- Delayed-onset pancreatitis secondary to migration of a fully covered self-expanding metal biliary stent: a case report.Journal of surgical case reports · 2026
- Staged Endoscopic Extraction of a Chronically Impacted Biliary Plastic Stent Using Temporary Parallel Decompression in a Resource-Limited Setting: A Case Report.International medical case reports journal · 2026