Gastrointestinal endoscopy

Comparative Effectiveness of Two Novel Anti-Migration Strategies in Fully Covered Self-Expandable Metal Stents for Biliary Strictures: A Multicenter Propensity-Matched Study.

Daryl Ramai, Andrew Fuller, Adam Bria, Khanh N C Duong, Ravi Pasam, Abdulrahman Qatomah, Bharat Ponnaganti, Pushpak Taunk,

Division of Gastroenterology, Hepatology, and Nutrition, University of Utah, Salt Lake City, UT, USA. Electronic address: Daryl.Ramai@hsc.utah.edu. · Division of Gastroenterology, Hepatology, and Nutrition, University of Utah, Salt Lake City, UT, USA.

Observational · n=228

Key points

From the paper's conclusion
  • FCSEMS-AB was associated with significantly higher migration risk compared with FCSEMS-AF, while demonstrating lower rates of stent occlusion. These findings highlight a trade-off between migration resistance and stent patency that may guide individualized stent selection.

Abstract

BACKGROUND AND AIMS: Fully covered self-expandable metal stents (FCSEMS) are used for benign and malignant biliary strictures, though stent migration and occlusion remain important limitations. A novel FCSEMS with an anchoring bump and enlarged distal flare (FCSEMS-AB) was recently introduced as an alternative anti-migration design to FCSEMS with anchoring fins (FCSEMS-AF). We compared migration and occlusion rates between these stent types. METHODS: We performed a multicenter retrospective cohort study of patients undergoing FCSEMS placement at three tertiary centers. Patients received either FCSEMS-AF (Viabil) or FCSEMS-AB (WallFlex Plus). Propensity score matching was performed to balance baseline characteristics. Logistic regression and Cox proportional hazards models were used to evaluate outcomes. RESULTS: Among 228 patients, 143 received FCSEMS-AF and 85 received FCSEMS-AB. After propensity matching, 64 matched pairs were analyzed. FCSEMS-AB was associated with significantly higher odds of stent migration compared with FCSEMS-AF in both unadjusted (OR 5.00, 95% CI 1.57-15.94, p=0.007) and adjusted analyses (aOR 5.25, 95% CI 1.58-17.49, p=0.007). Time-to-event analyses similarly demonstrated increased migration risk with FCSEMS-AB (adjusted HR 6.26, 95% CI 2.01-19.55, p=0.002). FCSEMS-AB was associated with lower odds of stent occlusion in adjusted logistic regression (aOR 0.27, 95% CI 0.10-0.78, p=0.015), though this association was not statistically significant in time-to-event analysis. Mortality did not differ between groups. CONCLUSIONS: FCSEMS-AB was associated with significantly higher migration risk compared with FCSEMS-AF, while demonstrating lower rates of stent occlusion. These findings highlight a trade-off between migration resistance and stent patency that may guide individualized stent selection.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers