Comparison of Testosterone in Umbilical Cord Blood of Preterm Infants Measured by Immunoassay and Mass Spectrometry.
Satomi Tanabe, Nobuhiko Nagano, Hiroki Terada, Yusuke Mine, Masako Aoki, Tetsuya Taniguchi, Junichi Suzuki, Mika Ishige,
Key points
From the paper's conclusion- In preterm infants at birth, immunoassays (ECLIA) demonstrated limited accuracy in measuring serum testosterone levels. Therefore, we recommend using LC-MS/MS because of its superior precision, clearer sex differentiation, and greater clinical utility.
Abstract
BACKGROUND: Testosterone plays an important role in male sexual differentiation. In adult males, electrochemiluminescence immunoassays (ECLIA) are used globally to measure serum testosterone levels because of their efficiency and good correlation with liquid chromatography-tandem mass spectrometry (LC-MS/MS), despite their cross-reactivity potential. However, the accuracy and clinical significance of these methods in the early neonatal period remain unclear. This study compared umbilical cord testosterone levels measured using ECLIA and LC-MS/MS to discriminate sex differences between two methods in preterm infants with typical external genitalia. METHODS: Forty-eight preterm infants admitted to the neonatal intensive care unit of the Nihon University Itabashi Hospital between May 2020 and March 2024 were recruited. Umbilical cord serum testosterone levels were measured using ECLIA and LC-MS/MS. Sex differences in testosterone levels between males and females were compared using the Mann-Whitney U test for median values and interquartile ranges (IQRs) (p-value < 0.05). The Spearman's correlation analysis, a non-parametric method, was performed to evaluate the correlations between testosterone levels measured using ECLIA, and testosterone levels measured using LC-MS/MS. Method comparison between ECLIA and LC-MS/MS was performed using Passing-Bablok regression. Logistic regression and receiver operating characteristic (ROC) curve analyses were performed to assess the ability of testosterone levels measured using ECLIA and LC-MS/MS to predict sex of preterm infants with typical external genitalia, respectively. RESULTS: This study included 38 preterm infants with typical external genitalia (16 males and 22 females). Testosterone levels measured using ECLIA were significantly higher and exhibited wider distributions (males: 1.10 ng/mL, IQR 0.90 - 2.69; females: 0.95 ng/mL, IQR 0.71 - 1.18) than those measured using LC-MS/MS (males: 0.210 ng/mL, IQR 0.164 - 0.327; females: 0.026 ng/mL, IQR 0.0112 - 0.0526) (p-value < 0.05). The correlation between testosterone levels measured using ECLIA and those measured using LC-MS/MS was weak (Spearman's rank correlation coefficient ρ(rho) < 0.5). In all infants, the Passing-Bablok regression yielded a slope of 0.082 and an intercept of -0.021 (Figure 3A), indicating limited agreement. Among preterm infants with typical external genitalia, the ROC curve analyses showed that LC-MS/MS had superior diagnostic accuracy than ECLIA, clearly differentiating males from females (area under the curve = 0.96 vs. 0.715, respectively). CONCLUSIONS: In preterm infants at birth, immunoassays (ECLIA) demonstrated limited accuracy in measuring serum testosterone levels. Therefore, we recommend using LC-MS/MS because of its superior precision, clearer sex differentiation, and greater clinical utility.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluation of a mass spectrometry-based method for estradiol determination in girls with benign premature thelarche - a method comparison with extraction RIA.Hormone research in paediatrics · 2026
- Assay-dependent reclassification of biochemical hyperandrogenism in female androgen testing: CLIA versus LC-MS/MS under common clinical decision thresholds.Clinica chimica acta; international journal of clinical chemistry · 2026
- Analytical performance and clinical classification value of a novel sandwich immunoassay for total testosterone measurement in low-concentration and hemodialysis populations.Clinical chemistry and laboratory medicine · 2026
- Efficacy of Liquid-chromatography and Radioimmunoassay in False-positives’ Drop-off in CAH Newborn ScreeningThe Journal of Clinical Endocrinology & Metabolism · 2026
- Preliminary analytical evaluation of the fully automated Cobas i601 LC-MS/MS system for steroid hormones and vitamin D in a clinical laboratory.Clinica chimica acta; international journal of clinical chemistry · 2026
- A-160 Estradiol Measurement in Patients Receiving Exogenous Estradiol Differs by Route of Administration and Immunoassay PlatformClinical Chemistry · 2026