ProbeFit: one-seizure qualification and personalisation of multimodal wearable seizure detectors
H. Cui, Y. Zhu, S. Huang, T. Xu
Key points
From the paper's conclusion- We propose ProbeFit to predict later seizure sensitivity and personalise a wearable detector from the first recorded seizure.
Abstract
Background and Objective: Missed seizures hinder assessment of seizure frequency and treatment. Although wearables can detect seizures automatically, performance varies between patients. Before deployment, it remains difficult to identify suitable patients, and personalisation usually requires several recorded seizures. We propose ProbeFit to predict later seizure sensitivity and personalise a wearable detector from the first recorded seizure. Methods: We used multicentre SeizeIT2 recordings of behind-the-ear EEG, ECG, EMG and movement from 125 patients. Two generic detectors' peak responses to the first seizure generated a qualification probability. The same seizure and preceding background personalised the detector. ProbeFit was developed on 96 patients and applied unchanged to 29 held-out patients. Expert-labelled video-EEG seizures provided the reference. Evaluation compared qualification with the first-seizure heart-rate rule and measured discrimination, calibration and net benefit, and it compared later sensitivity of the generic and personalised detectors at equal false-alarm rates. Results: Among 27 held-out patients with 152 later seizures, qualification correlated with later sensitivity at Spearman {rho} = 0.38, with one-sided p = 0.033. It identified patients with later sensitivity [≥] 0.5 with an area under the ROC curve (AUC) of 0.74. Across all 97 evaluable patients, AUC was 0.76 against 0.59 for the heart-rate rule, a difference of 0.17 with a 95% confidence interval of 0.05 to 0.29. At the qualification threshold of 0.5, ProbeFit withheld deployment from 9 of 15 held-out patients below the sensitivity criterion, and its net benefit was 0.074 against -0.111 for selecting everyone. At equal false-alarm rates between 0.25 and 2 per day, personalisation raised later sensitivity by 2.2 percentage points in held-out patients and by 6.3 points in development patients. Both confidence intervals excluded zero, and no held-out patient lost sensitivity. Conclusions: ProbeFit uses the first recorded seizure to predict whether a wearable detector will detect at least half of a patient's later seizures, and it personalises that detector without waiting for further seizures. It supports patient selection and adaptation during the first monitoring admission.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Seizure detection using wearable electrocardiogram connected to a smartphone: a phase 3 clinical validation studyEBioMedicine · 2025
- Seizure quantification using wearable devices.Epilepsia · 2026
- Beyond sensitivity: Performance, methodological heterogeneity, and real-time translation of wrist-worn convulsive seizure detection devices - A systematic review and meta-analysis.Seizure · 2026
- Wearable Artificial Intelligence for Epilepsy: Scoping ReviewJournal of Medical Internet Research · 2025
- Modalities and algorithms for generalized motor seizure detection and prediction: a scoping reviewExpert Review of Medical Devices · 2026
- Seizure Prediction Using Channel-Optimized Intracranial EEG With Variable Interictal and Preictal Durations.IEEE open journal of engineering in medicine and biology · 2026