Beyond sensitivity: Performance, methodological heterogeneity, and real-time translation of wrist-worn convulsive seizure detection devices - A systematic review and meta-analysis.
Mohammed A Fouda, Cameryn A Freglette, H Allison Bender, Megan Still, Parsa Nilchian, Caitlin E Hoffman
Department of Neurological Surgery, New York-Presbyterian Hospital, Weill Cornell Medicine, NY, USA. Electronic address: Maf4023@med.cornell.edu. · Department of Neurological Surgery, New York-Presbyterian Hospital, Weill Cornell Medicine, NY, USA.
Key points
From the paper's conclusion- Wrist-worn devices demonstrate high sensitivity for convulsive and GTC seizure detection, but performance is not yet methodologically uniform or consistently generalizable.
Abstract
Wrist-worn seizure-detection devices can detect convulsive seizures, but pooled performance estimates may obscure methodological differences between studies. We evaluated the performance, heterogeneity, and translational readiness of wrist-worn non-EEG systems for automated detection of convulsive and generalized tonic-clonic (GTC) seizures. We systematically searched MEDLINE/PubMed, Embase, Scopus, Web of Science, and Google Scholar. Event-level sensitivity for convulsive and GTC seizures was pooled using random-effects generalized linear mixed models; false-alarm rates (FARs) were pooled as incidence rates per 24 h. We evaluated heterogeneity, prespecified subgroups, methodological variability, detection latency, and progression to real-time intervention. Twenty studies comprising 1598 participants and 71,310 monitoring hours were included. Pooled random-effects sensitivity was approximately 93% for convulsive seizures and 96% for GTC seizures, with heterogeneity exceeding 75% for both outcomes. The pooled FAR was approximately 0.39/24 h, with heterogeneity approaching 99%. Neither sensing modality, study phase, population, algorithm locking, nor offline versus real-time processing explained performance differences under random-effects modeling. Methodological variability extended to seizure eligibility, amplitude and duration thresholds, temporal windowing, classifier architecture, operating-point selection, validation strategy, and FAR denominator construction. Mean detection latency was 30.04 s across nine reporting studies. Only one study directly evaluated translation from detection to real-time caregiver notification and intervention. Wrist-worn devices demonstrate high sensitivity for convulsive and GTC seizure detection, but performance is not yet methodologically uniform or consistently generalizable. Standardized evaluation and independent validation are needed, with future studies extending beyond detection accuracy to reliable real-time alert delivery and actionable intervention.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Seizure quantification using wearable devices.Epilepsia · 2026
- Modalities and algorithms for generalized motor seizure detection and prediction: a scoping reviewExpert Review of Medical Devices · 2026
- Seizure detection using wearable electrocardiogram connected to a smartphone: a phase 3 clinical validation studyEBioMedicine · 2025
- ProbeFit: one-seizure qualification and personalisation of multimodal wearable seizure detectorsmedRxiv · 2026
- Wearable Artificial Intelligence for Epilepsy: Scoping ReviewJournal of Medical Internet Research · 2025
- Selection of Low-Density EEG Montages for Optimal Seizure Detection Performance.Healthcare technology letters · 2026