Circular-Arc Modal Migration for Ultrasound Imaging--Part II: Simulation and Experimental Evaluation in Compact Semicircular and Convex Arrays
Z. Huang, Z. Liu, H. Qi, C. Chen, D. Wu
Key points
From the paper's conclusion- At 512 x 513 pixels and 24 transmissions, median host latency is 29.79 ms for DAS and 51.87 ms for CAMM on an RTX 4090.
Abstract
Circular-arc ultrasound imaging requires reconstruction that accounts for curvature and offset transmit sources across a wide field. This study evaluates how Circular-Arc Modal Migration (CAMM) changes point-response width, image-energy distribution, and low-echo target separation relative to delay-and-sum (DAS). CAMM combines angular modal analysis, local spectral remapping, and radial synthesis, with the offset phase evaluated through phase screens or approximated by scalar pullback. Guided by Part I's error analysis and point-target comparisons, paired reconstructions use scalar pullback for compact semicircular cyst simulations and phase screens for acquisition-matched C5-2v simulations, six phantom records, and two carotid views from one participant. Each DAS/CAMM pair uses identical RF. The principal observation is that lower energy fractions outside a fixed neighborhood can coexist with broader tangential responses. In C5-2v point simulations, the full-field fraction is lower at 23 of 24 locations, whereas tangential width is greater at all 24. Semicircular cyst contrast improves in most configurations, with reversed ranking at the smallest target. All eight C5-2v cyst configurations at 10--15-mm radial depth favor CAMM in the three contrast measures; deeper comparisons depend on location and metric. Measured reflectors show the same width--energy-fraction pattern, and the evaluated phantom cyst and carotid lumen regions show improved contrast. At 512 x 513 pixels and 24 transmissions, median host latency is 29.79 ms for DAS and 51.87 ms for CAMM on an RTX 4090. The tested CAMM implementation therefore provides regional gains in low-echo target separation alongside broader tangential responses and higher computational cost.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Circular-Arc Modal Migration for Ultrasound Imaging--Part I: Finite-Arc Spectral Theory and Stolt-Pullback FactorizationbioRxiv · 2026
- A 240-Element Matrix Probe With Aberration Mask for 4-D Carotid Artery Computational Ultrasound Imaging.IEEE transactions on ultrasonics · 2026
- Microbeamformer-aware layout optimization of a 2D intracardiac echocardiography array for grating-lobe suppression.Ultrasonics · 2026
- Hadamard-Based Recursive Aperture Decoded Ultrasound Imaging (READI) With Estimated Motion-Compensated Compounding (EMC2) Using Top-Orthogonal-to-Bottom-Electrode (TOBE) Arrays.IEEE transactions on ultrasonics · 2026
- C2-ULM: Data-Efficient ultrasound localization microscopy via coherent prior MAP beamforming for Sparse, compressed receive acquisition.Ultrasonics · 2026
- Subwavelength registration of skull-compensating acoustic holograms via nonlinear parametric array feedbackarXiv · 2026