A single-plane electrical impedance tomography framework for integrated liver and kidney assessment
A. Wu, J.-C. Lai, M. K. Ho, J. Q. J. Ge, N.-j. Gong, I. Y. Zhou, M. Wong, B. J. Edelman,
Key points
From the paper's conclusion- Together, these findings support the feasibility of a portable multi-organ EIT imaging framework capable of extracting complementary liver and kidney information from a single belt placement.
Abstract
Chronic liver and kidney diseases frequently coexist, but their assessment typically requires separate clinical workflows involving serum-based estimated glomerular filtration rate (eGFR) testing and liver elastography-based controlled attenuation parameter (CAP). Electrical impedance tomography (EIT) has recently shown promise for providing portable, non-invasive proxies for both reference measurements. However, existing multi-organ EIT assessment still requires different acquisition protocols for each organ. To simplify this procedure, we assessed the feasibility of a single-plane EIT framework for reconstructing organ-specific images and estimating liver and kidney health from one abdominal belt placement. Across diverse abdominal belt locations, opposite-plane deviation was significantly greater for eGFREIT estimates (2.92%; IQR, 0.46-12.53%) than for CAPEIT estimates (0.94%; IQR, 0.66-1.70%) (p < 0.01). For both organs, compared to images acquired at the conventional organ-specific planes, image correlation significantly decreased (both, p < 0.05) and normalized root mean squared error significantly increased (both, p < 0.05) at the furthest imaging plane, with these differences being numerically larger for the kidney. Together, these results supported the kidney plane as the preferred location for multi-organ assessment. Test-retest evaluation after belt removal and reapplication demonstrated that the correlation between two repeated CAPEIT estimates was just as strong at the kidney plane (R2 = 0.98, p < 0.001) as it was at the conventional liver plane (R2 = 0.97, p < 0.001). Finally, in an independent clinical cohort, kidney-plane CAPEIT estimates were significantly higher in participants with a high metabolic liver-risk profile (i.e. diabetes and high blood pressure) (p < 0.005). CAPEIT estimates also distinguished high and low metabolic risk groups with an AUC of 0.756. Together, these findings support the feasibility of a portable multi-organ EIT imaging framework capable of extracting complementary liver and kidney information from a single belt placement.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- EIT reconstruction parameters’ impact on ventilation monitoring featuresCurrent Directions in Biomedical Engineering · 2026
- Data Archive for 2D and 3D Electrical Impedance Tomography with moderate conductivity contrast phantomsarXiv · 2026
- Gaussian decomposition of EIT reconstructions with different inclusion geometries.Physiological measurement · 2026
- Electrical Impedance Tomography Accurately Reflects Regional Ventilation Distribution Compared With Computed Tomography in Non-Intubated SubjectsRespiratory Care · 2026
- Recommendations for lung ventilation and perfusion assessment with chest electrical impedance tomography in critically ill adult patients: an international evidence-based and expert Delphi consensus studyEClinicalMedicine · 2025
- Electrical impedance tomography-guided lateral positioning nursing pathway in an older patient with severe ARDS unable to tolerate prone positioning: a case reportFrontiers in Medicine · 2026