Data Archive for 2D and 3D Electrical Impedance Tomography with moderate conductivity contrast phantoms
S. J. Hamilton, A. Hauptmann, A. H\"anninen, V. Kolehmainen, J. Niskanen, T. Savolainen, J. Toivanen
Key points
From the paper's conclusion- The data from thoracic imaging inspired cylindrical and chest-shaped phantoms are suitable for 2D algorithms, and the data from an intracerebral hemorrhage imaging inspired head-shaped phantom with resistive skull are suitable for 3D algorithms.
Abstract
This document describes open Electrical Impedance Tomography (EIT) measurement data that were collected from biomedical imaging inspired phantoms with moderate conductivity contrasts. The data were collected with the KIT4 and KIT5 EIT measurement devices at the University of Eastern Finland, and are intended for EIT image reconstruction algorithm development and testing. The data from thoracic imaging inspired cylindrical and chest-shaped phantoms are suitable for 2D algorithms, and the data from an intracerebral hemorrhage imaging inspired head-shaped phantom with resistive skull are suitable for 3D algorithms. In addition to the measurement data, computational meshes and example reconstruction codes are provided and sample reconstructions included.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A flexible finger-shaped tactile sensor based on electrical impedance tomographyJapanese Journal of Applied Physics · 2026
- Gaussian decomposition of EIT reconstructions with different inclusion geometries.Physiological measurement · 2026
- Feasibility of prior‐informed domain‐based electrical impedance tomography for deep tissue thermometry: Phantom experiments and computed tomography‐derived numerical simulationsMedical Physics · 2026
- Contrast-based detectability of structural gaps and hollow defects using 2D and 3D electrical impedance tomography: a finite element evaluationEngineering Research Express · 2026
- Implementing the EIT complete electrode model through Robin boundary conditionsarXiv · 2026
- A Joint Measurement and Sensitivity Enhancement Method for Improving EIT Image Reconstruction Quality.Sensors (Basel, Switzerland) · 2026