Virtual Noncontrast Imaging with Deep-Silicon Photon-Counting CT: Matched Phantom and Initial Human Evaluation

N. Azimi, A. Brian-Choux, S. Feemster, A. Rahbar, H. Linder, D. Crotty, Z. Yin, P. Balthazar,

Observational

Key points

From the paper's conclusion
  • To compare virtual noncontrast (VNC) error between deep-silicon photon-counting CT (PCCT) and dual-energy energy-integrating detector CT (EID-CT) in a matched phantom and characterize tissue-specific error in humans.

Abstract

To compare virtual noncontrast (VNC) error between deep-silicon photon-counting CT (PCCT) and dual-energy energy-integrating detector CT (EID-CT) in a matched phantom and characterize tissue-specific error in humans. This prospective exploratory study included a matched phantom experiment and pilot human evaluation. Prototype projection-domain water-density maps served as the VNC surrogate. Twenty-four vials containing 0.1--20 mg I/mL in deionized water or simulated blood were scanned with PCCT and EID-CT. Three men underwent true noncontrast and contrast-enhanced PCCT; 80 matched measurements were analyzed. Paired t tests and mixed-effects models assessed system differences and iodine associations. Three participants (mean age, 67 {+/-} 10 years; all men) were included. Mean VNC error was 29.1 HU lower with PCCT than EID-CT (95% CI, 15.5--42.7 HU; {rho} < 0.001). Error was not associated with iodine concentration for PCCT (slope, 0.03 HU per mg I/mL; 95% CI, -0.92 to 0.98; {rho} = 0.95) but increased for EID-CT (slope, 3.51 HU per mg I/mL; 95% CI, 2.56--4.46 HU; {rho} < 0.001). The slopes differed by 3.48 HU per mg I/mL ({rho} < 0.001). In humans, errors in five non-blood-pool tissues were within 5 HU. Error was not associated with iodine concentration among 61 non-blood-pool measurements (adjusted slope, -0.17 HU per mg I/mL; 95% CI, -1.22 to 0.88; {rho} = 0.75). PCCT provided VNC measurements with errors closer to zero and less dependent on iodine than EID-CT in the phantom. In humans, error was generally small in non-blood-pool tissues and was not associated with iodine concentration.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Virtual Noncontrast Imaging with Deep-Silicon Photon-Counting CT: Matched Phantom and Initial Human Evaluation — Science Skim