IMU-based upper limb kinematics validation: sensor fusion optimisation, sensor-to-segment alignment, and joint angle accuracy.
A. Fabre, G. Kerr, P. Pivonka, W. van den Hoorn
Key points
From the paper's conclusion- These results indicate that IMU-derived kinematic accuracy is context-dependent, varying with joint and movement type, suggesting that accuracy should be established for the specific movements and joints relevant to each clinical application.
Abstract
Conventional clinical assessment of upper limb kinematics relies on goniometry or visual estimation, methods that provide limited information on multi-segment 3D kinematics and can have poor inter-rater reliability. Inertial measurement units (IMUs) represent a viable alternative; however, extracting anatomically meaningful joint angles from skin-mounted sensors remains challenging owing to sensor-to-segment misalignment, soft-tissue artefact, and suboptimal orientation estimation under dynamic conditions. This study developed and validated a multi-segment IMU-based framework for upper limb kinematic assessment, addressing three stages of the measurement pipeline: (i) sensor fusion gain optimisation as a function of task mean segment angular velocity; (ii) accuracy of an anatomical landmark registration-based alignment method; and (iii) comparative evaluation of landmark-based alignment against manual and neutral-pose alignment approaches. Twenty participants with typical shoulder function performed clinical and functional upper limb tasks, recorded simultaneously with an optoelectronic motion capture system and six IMUs tracking the thorax, clavicle, scapula, humerus, forearm, and hand. Orientation estimation used a gradient-descent complementary filter with velocity-dependent gain modulation. Sensor-to-segment alignment was achieved via manual, neutral-pose, and landmark registration methods. Mean sensor orientation errors ranged from 1.5 to 3 degrees across segments/tasks and increased with angular velocity. Landmark registration produced substantially lower joint angle errors than manual and neutral-pose alignment across all joints (root mean square error range: 3.7-7.8 degrees; manual: 8.3-24.2 degrees; neutral: 15.5-35.5 degrees). These results indicate that IMU-derived kinematic accuracy is context-dependent, varying with joint and movement type, suggesting that accuracy should be established for the specific movements and joints relevant to each clinical application.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Inertial human motion capture: From biomechanics to recent sensor fusion methods and back.Journal of biomechanics · 2026
- IMU- and Vision-Based Measurement Techniques for Joint Kinematics: A Narrative Review.Sensors (Basel, Switzerland) · 2026
- A Motion-Capture-Based Method for Upper-Limb Joint-Angle EstimationEngineering Research Express · 2026
- Agreement of Wearable Inertial Sensors with Optoelectronic Motion Capture for Felge-to-Support Temporal Segmentation.Bioengineering (Basel, Switzerland) · 2026
- Criterion Validity of Wrist-Worn IMU-Derived Smoothness Metrics During Functional Upper-Limb Tasks.Bioengineering (Basel, Switzerland) · 2026
- Measurement Uncertainty and Traceability in Upper Limb Rehabilitation Robotics: A Metrology-Oriented ReviewJournal of Sensor and Actuator Networks · 2026