Smartphone-connected, battery-free aptamer-based digital test for quantification of procalcitonin and monitoring of infection at the point of care
Yuanjun Cai, José M.R. Flauzino, Abdulkadir Sanli, Tinghao Hu, Hong S. Lee, Alex S. Collins, Laura Gonzalez‐Macia, Suzanne Williams,
Imperial College London · Fleming College
Key points
From the paper's conclusion- We developed a battery-free, smartphone-connected electrochemical lateral flow assay with linked analytics for the detection of PCT (ELLA-PCT) enabling quantitative testing at the point of care without conventional laboratory instrumentation.
Abstract
Summary Background Rapid and accurate measurement of procalcitonin (PCT) is useful for diagnosing bacterial infections and guiding antibiotic therapy, yet current laboratory-based immunoassays require centralised infrastructure, delaying clinical decision-making and limiting access in low- resource settings. We developed a battery-free, smartphone-connected electrochemical lateral flow assay with linked analytics for the detection of PCT (ELLA-PCT) enabling quantitative testing at the point of care without conventional laboratory instrumentation. Methods We designed a competitive electrochemical lateral flow assay using gold nanoparticles co-functionalised with a PCT-specific DNA aptamer and ferrocene hexanethiol as a redox reporter. A miniaturised near-field communication (NFC) potentiostat embedded within a disposable cassette enabled wireless electroanalytical measurements using a smartphone. Analytical performance was assessed in buffer and serum, including limit of detection (LOD), linearity, specificity, and stability. Clinical evaluation was performed on 27 serum samples from nine adults undergoing antibiotic treatment for suspected bacterial infection, with results compared against the reference Time-Resolved Amplified Cryptate Emission (TRACE) assay. Findings ELLA-PCT achieved an LOD of 46 pg/mL and a linear detection range of 0.5–100 ng/mL. The ELLA-PCT assays remained stable for three months under ambient storage conditions, and cross-reactivity with calcitonin, C-reactive protein, and interleukin-6 remained below clinically relevant thresholds. Clinical results showed strong correlation with TRACE (R² = 0·979, p < 0·0001), with a mean bias of 0·05 ng/mL and narrow 95% limits of agreement (– 0·69 to 0·89 ng/mL). Importantly, ELLA-PCT delivered results in only 30 minutes without requiring external power or clinical laboratory instrumentation. In comparison to TRACE, the total turnaround time was reduced by at least 50%, which typically requires at least one hour from sample collection to results at a hospital setting. Interpretation ELLA-PCT is an antibody-free, smartphone-connected test that provides laboratory-grade quantitative PCT measurement using only an NFC-enabled electroanalytical sensor and a mobile phone. The platform has a strong potential to decentralise diagnosis of infectious diseases, support antibiotic stewardship, and enable remote, real-time monitoring in outpatient and resource-limited settings. Larger studies are needed to evaluate the use of whole- blood samples and the integration of the platform into digital clinical workflows. Funding The authors acknowledge the Gates Foundation (Grand Challenges Explorations scheme under grant numbers OPP1212574 and INV-038695) for financial support. FG and JMRF acknowledge Research England International Science Partnerships Fund (References: G08259 and G08288) and the Biotechnology and Biological Sciences Research Council Impact Acceleration Account (Reference: PSR417). FG and AS acknowledge Innovate UK BMC (Reference: PA2131). FG and LGM acknowledge funding from the European Union’s Horizon 2020 research and innovation programme under the Marie Sklodowska-Curie grant agreement No 101025390. FG acknowledge the Bezos Earth Fund through the Bezos Centre for Sustainable Protein (BCSP/IC/001). TMR, RW, SF, & SW acknowledge The Wellcome Trust funded Centres for Antimicrobial Optimisation Network (CAMO-NET. Reference: 226691/Z/22/Z)
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Antibody- and enzyme-free point-of-care detection of procalcitonin via synergistic nanomachines-activated catalytic assembly (SNCA) assay for early sepsis monitoring.Biosensors & bioelectronics · 2026
- Smartphone-Controlled Point-of-Care Testing Platform Integrating FTA Card-Based Sample Preparation and Colorimetric LAMP for Pathogen DetectionMicromachines · 2026
- Development and application of a smartphone-based thermal contrast amplification reader for photothermal lateral flow immunoassays.Microsystems & nanoengineering · 2026
- Smartphone-assisted upconversion nanoparticles-aptamer-based probe functionalized hydrogel microneedles for rapid interstitial fluid lactate profiling.Talanta · 2026
- Spherical Ag2O nanoparticles with laccase-mimicking activity: a smartphone-assisted and label-free colorimetric aptasensor for Pseudomonas aeruginosa detection in drinking water.Talanta · 2026
- Disposable laser-induced graphene-based biosensor strip for the detection of N-protein using CRISPR activation and ratiometric electrochemical readoutTalanta · 2026