Comparison of different solutions for applying patient-specific beam quality in treatment planning CT imaging.
Ákos Sudár, Tamás Pócza, Domonkos Szegedi, Tibor Major, Ádám Domonkos Tárnoki, Dávid László Tárnoki, Csilla Pesznyák
Oncologic Imaging and Invasive Diagnostic Centre, National Institute of Oncology, Budapest, Hungary. · Center of Radiotherapy, National Institute of Oncology, Budapest, Hungary.
Key points
From the paper's conclusion- A traditional kernel with calibration curves for different energies results in more accurate electron density determination than the Artificial 120 kVp or the Direct Density reconstruction. One of the most widely used treatment planning systems has recently introduced the option to add multiple calibration curves for the same CT model.
Abstract
BACKGROUND: For radiotherapy treatment planning, it is necessary to convert Hounsfield units from CT images to electron density. This conversion can be performed using separate calibration curves for each beam quality. Another available solution is to use an Artificial 120 kVp reconstruction, allowing the use of a single calibration curve. The Direct Density kernel is a more modern approach that directly reconstructs electron density and requires no calibration measurements. MATERIALS AND METHODS: The Gammex RMI 467 phantom was measured at different positions: the central axis and an anterior offset from it. The patient-size dependence was tested by measuring the Gammex inserts in two in-house-made phantoms with variable water-equivalent diameters. RESULTS: It was found that the standard reconstruction with separate calibration curves is the most accurate solution. The Artificial 120 kVp reconstruction introduces slightly greater uncertainty in the reconstructed electron density (ED), increasing the tube potential dependence of the reconstructed ED. The Direct Density reconstruction is less accurate; however, its performance can be improved by calibration. CONCLUSIONS: A traditional kernel with calibration curves for different energies results in more accurate electron density determination than the Artificial 120 kVp or the Direct Density reconstruction. One of the most widely used treatment planning systems has recently introduced the option to add multiple calibration curves for the same CT model.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- CT-ED/MD calibration curve commissioning and end-to-end validation of HyperSight CBCT-Based dose calculation for online adaptive radiotherapy.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Technical note: reconstructing dose distributions from manually planned electron boosts in breast radiotherapy.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Quantitative comparison of 68-keV monoenergetic, routine-dose, and low-dose 120-kVp CT for calcium-mimicking inserts in thoracic phantoms: effects of phantom size and reconstruction method.European radiology experimental · 2026
- Comparison of three commercial multi-energy CT and electron density phantoms for iodine quantification and tissue substitute characterization.Journal of applied clinical medical physics · 2026
- Influence of phantom size on the IEC 60601-2-44 wide-beam CTDI correction: A Monte Carlo studyEPJ Web of Conferences · 2026
- Technical assessment of a novel vertical CT system for upright radiotherapy simulation and treatment planningMedical Physics · 2026