VNA-Only Calibration of Waveguide Thermistor Mounts with Self-Consistent Determination of the Thermistor-Port Impedance
Patricio Greco
Key points
From the paper's conclusion- The scattering matrix implied by the solution is not passive; a minimum-norm correction of the thermistor impedances along the two directions invisible to the calibration factor - the ratios of the efficiencies of the three states - brings it to the floor of passivity, where the efficiency depends on the load only through its mismatch factor, leaves the result unchanged, and gives the sensitivity
Abstract
A method is presented for determining the calibration factor and the effective efficiency of waveguide thermistor mounts from vector-network-analyzer reflection data alone, without a microcalorimeter, extending the three reflection-coefficient method of the coaxial case. In waveguide the thermistor is coupled to the line through a matching structure and the reference impedance of the thermistor port, $Z_0$, is not known a priori; it is inferred through a power-consistency relation between pairs of states, under the convention that it is real. The method reproduces the factory certificates (U = 2 %) of four K486A mounts within 1.0 % rms and the PTB-traceable values (U = 0.85 to 0.98 %) of a fifth within 0.34 % rms; its expanded measurement uncertainty is 0.5 to 1.16 % (k = 2) for the best specimen and below 2 % for the five. A mount with its waveguide-to-coaxial adapter is calibrated directly, without characterizing the adapter, from the model factor of the mount and the geometric factor measured through the adapter. The scattering matrix implied by the solution is not passive; a minimum-norm correction of the thermistor impedances along the two directions invisible to the calibration factor - the ratios of the efficiencies of the three states - brings it to the floor of passivity, where the efficiency depends on the load only through its mismatch factor, leaves the result unchanged, and gives the sensitivity of the efficiency to the bias resistance in closed form: below $4\times10^{-5}$ $\Omega^{-1}$ in band mean, proportional to the mismatch of the operating state and to the loss of the mount.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Passivity Correction of the Thermistor Model in Bolometric Power Sensors and Its Dependence on Bead TemperaturearXiv · 2026
- Review on the Precise Microwave Brightness Temperature Transfer in the Radiometer Calibration LinkRemote Sensing · 2026
- Development and validation of a multichannel edge density profile reflectometer embedded in ion cyclotron resonance heating (ICRH) antennas on EAST.The Review of scientific instruments · 2026
- Uncertainty Analysis and Metrological Validation of Raman Distributed Temperature Measurements in a Full-Scale Test FacilitySensors · 2026
- Waveguide-based measurement of the silicon nitride refractive index and thermo-optic coefficient from the visible to the near-infraredAPL Photonics · 2026
- Low loss PTFE core metal/dielectric waveguide for W band applicationsInternational Journal of Microwave and Wireless Technologies · 2026