Comprehensive clinical dosimetry in contact skin interventional radiotherapy (modern brachytherapy): Practical tips and tricks.
Bruno Fionda, Enrico Rosa, Elisa Placidi, Valentina Lancellotta, Pierpaolo Dragonetti, Maria Concetta La Milia, Martina De Angeli, Francesco Pastore,
UOC Degenze di Radioterapia Oncologica, Dipartimento di Diagnostica per Immagini e Radioterapia Oncologica, Fondazione Policlinico Universitario A. Gemelli IRCCS, Rome, Italy. · UOC Fisica per le Scienze della Vita, Dipartimento di Diagnostica per Immagini e Radioterapia Oncologica, Fondazione Policlinico Universitario A. Gemelli IRCCS, Rome, Italy.
Key points
From the paper's conclusion- Although clinically non-invasive, cIRT requires careful dosimetric optimization. Appropriate management of applicator geometry, dose calculation, bolus, air gaps, and patient-specific devices, may improve treatment accuracy, target coverage, and toxicity control.
Abstract
PURPOSE: Contact interventional radiotherapy (cIRT) is an effective option for non-melanoma skin cancers, particularly in anatomically complex sites, where organ preservation, functional results, and cosmetic outcomes are relevant. This paper summarized practical dosimetric considerations in cIRT, focusing on the therapeutic window, surface curvature, boluses and air gaps, dose calculation algorithms, and 3D-printed devices. MATERIAL AND METHODS: Key physical and clinical factors affecting dose distribution were analyzed from recent literature, including prescription depth, source-to-skin distance, applicator geometry, tissue interfaces, and patient-specific device design. RESULTS: Therapeutic window is influenced by target thickness, source-to-skin distance, and surface curvature. Multilayer and intensity-modulated arrangements may improve target coverage while limiting superficial hotspots. Bolus and air gaps can modify local dose distribution, especially in irregular anatomical regions, while model-based algorithms better account for heterogeneities and interfaces than TG-43. 3D-printed applicators may further enhance conformity, reproducibility, and organs at risk sparing. CONCLUSIONS: Although clinically non-invasive, cIRT requires careful dosimetric optimization. Appropriate management of applicator geometry, dose calculation, bolus, air gaps, and patient-specific devices, may improve treatment accuracy, target coverage, and toxicity control.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Impact of applicator curvature in contact interventional radiotherapy (brachytherapy) for skin cancer: exploring the “Nautilus Effect”Physica Medica · 2026
- Minimum recommended distances between cardiac implantable electronic devices and the applicator surface in low-energy intraoperative radiotherapy of the breast: an in vivo dosimetric study.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Development and Preliminary Dosimetric Evaluation of a 3D-Printed Custom Tissue Compensator for Head and Neck Cancer Radiotherapy: A Proof-of-Concept Case Series.Technology in cancer research & treatment · 2026
- Updated dosimetric comparison of HDR interventional radiotherapy including 60Co sources versus VMAT and electron beam therapy for NMSC.Radiation and environmental biophysics · 2026
- Modern brachytherapy modalities vs. external beam radiotherapy in post-operative non-melanoma skin cancer: A systematic review.Journal of contemporary brachytherapy · 2026
- A 3D CBCT-Based Approach to Assessing Effective Radiation Penetration Depth in Intraoperative Electron Radiotherapy.Diagnostics (Basel, Switzerland) · 2026