Shockingly Unreliable: A Case Report of Failure of Synchronized Cardioversion Device to Capture, Resulting in Ventricular Fibrillation and Cardiac Arrest.
Clifford J Buckley, Bradley J McClintic, Andrew L Juergens, Shea Standley
Baylor Scott & White Medical Center, Department of Emergency Medicine, Temple, TX.
Key points
From the paper's conclusion- We present the case of a 65-year-old woman with unstable wide-complex tachycardia with a pulse who underwent emergent synchronized cardioversion.
Abstract
UNLABELLED: Synchronized cardioversion is a standard treatment for unstable ventricular tachycardia. While modern defibrillators are designed with safety mechanisms, synchronization failure can result in catastrophic complications. We present the case of a 65-year-old woman with unstable wide-complex tachycardia with a pulse who underwent emergent synchronized cardioversion. Despite activation of the "sync" mode, the device failed to capture appropriately and delivered a shock on the T-wave, precipitating ventricular fibrillation and cardiac arrest. After 26 minutes of resuscitation and multiple defibrillations, return of spontaneous circulation (ROSC) was achieved. The patient ultimately recovered and was discharged home in stable condition. This case underscores the rare but serious risk of synchronization failure, particularly in obese patients where pad placement and capture may be impaired,1 and emphasizes the importance of visually confirming synchronization prior to shock delivery. TOPICS: Cardioversion, ventricular fibrillation, defibrillator malfunction, ACLS, emergency medicine.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Loss of Biventricular Capture After CRT-D Implantation: Device Malfunction or Something Else?JACC. Case reports · 2026
- Vital sign parameters are not associated with electrical capture during transcutaneous pacing.The American journal of emergency medicine · 2026
- Occult ventricular fibrillation during on-pump beating coronary artery bypass grafting: A case reportJCA advances. · 2026
- Emergency Transcatheter Aortic Valve Implantation in Mixed Bicuspid Aortic Valve Disease With a Horizontal Aorta: Return of Spontaneous Circulation Only After Left Ventricular Unloading.The American journal of cardiology · 2026
- Failure in executing the defibrillation threshold test after subcutaneous implantable cardioverter defibrillator implantation as a predictor of adverse clinical outcomesEuropean Heart Journal · 2026
- Does transthoracic impedance influence termination and recurrence of ventricular fibrillation during impedance-compensated shocks for out-of-hospital cardiac arrest? A retrospective studyResuscitation · 2026