Development, Validation, and Application of a Highly Sensitive RP-HPLC Method for the Determination and Release Profile Comparison of Levonorgestrel from Silicone-Based Contraceptive Implants.
Abul Kalam Lutful Kabir, Sitesh Chandra Bachar, Md Moniruzzaman, Emdad Hossain
University of Dhaka, Department of Pharmaceutical Technology, Dhaka, Bangladesh. · Techno Drugs Ltd., Dhaka, Bangladesh.
Key points
From the paper's conclusion- The devolved method was straightforward and quick for any sample preparation, which could be used for routine analysis, in-process quality control analysis, and stability study analysis, among others. Moreover, Femiplant presented similar release kinetics compared to the World Health Organization prequalified implant Sino-implant (II), thereby having similar bioavailability and ensuring the consistent therapeutic effect of both implants.
Abstract
OBJECTIVES: Levonorgestrel (LNG), a synthetic progestin containing implants, has emerged as one of the most reliable, effective, and long-acting reversible contraceptive (LARC) methods for preventing unintended pregnancies. The current study aimed to develop and validate a simple, rapid, and highly sensitive reverse-phase high-performance liquid chromatography (HPLC) technique for the quantification and release profile comparison of LNG from different silicone-based LARC implants. MATERIALS AND METHODS: All the analysis was performed using HPLC equipped with UV/UV-VIS detector with the help of Waters C18 column of 250 × 4.6 mm dimension, 5 μm sized particle with mobile phase consisting of water and acetonitrile in the ratio of 30:70 (v/v%) at room temperature and 1 mL/minute flow rate, injection volume of 50 μl with the detection wavelength of 240 nm. The developed method was then validated as per International Council for Harmonisation of Technical Requirements for Pharmaceuticals for Human Use (ICH) guidelines and then the validated method was used successfully to quantify LNG and for a comparative release profile study of two marketed implants, Femiplant and Sino-implant (II). The obtained release data were assessed and compared using fit factors (f1 & f2) to compare one formulation with another. RESULTS: All parameters met the limits according to ICH guidelines. The total time needed for analysis was 12 minutes, having LNG eluting with a retention time of 5.8 min (± 0.2). The average LNG content was obtained to be 98.52 to 101.32% (± 0.05%). From the release kinetic study, f1 and f2 values of Femiplant against Sino-implant (II) determined was 10 and 62, respectively, both of which meet the acceptable limits. CONCLUSION: The devolved method was straightforward and quick for any sample preparation, which could be used for routine analysis, in-process quality control analysis, and stability study analysis, among others. Moreover, Femiplant presented similar release kinetics compared to the World Health Organization prequalified implant Sino-implant (II), thereby having similar bioavailability and ensuring the consistent therapeutic effect of both implants.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Green and lean: A validated RP-HPLC-VWD method for simultaneous determination of hydroquinone, methylparaben, fluocinolone acetonide, and propylparaben.Journal of chromatography. B, Analytical technologies in the biomedical and life sciences · 2026
- Development and Validation of an RP‐HPLC Method for the Rapid and Sensitive Quantification of Everolimus in Pharmaceutical and Medical Device ApplicationsSEPARATION SCIENCE PLUS · 2026
- Pharmacokinetic bioequivalence between (etonogestrel/ethinylestradiol 11.7/2.7 mg) NuvaRing® and ornibel premium® (etonogestrel/ethinylestradiol 8.25/2.6 mg), a novel polymer composition contraceptive vaginal ring.The European journal of contraception & reproductive health care : the official journal of the European Society of Contraception · 2026
- Development and validation of a LC-MS/MS method for the quantification of the E-type prostanoid receptor 4 antagonist HTL0039732 in human plasma, for a Phase I/IIa clinical trial.Bioanalysis · 2026
- Development and Validation of an Ultra‐Fast UPLC–MS/MS Method for Simultaneous Quantification of Lodenafil and Lodenafil Carbonate in Human Liver Microsomes: Depletion Rate and Metabolic Stability AssessmentBiomedical Chromatography · 2026
- Development and Preclinical Evaluation of a Dual-Drug Implant for Long-Acting HIV Prevention and Contraception.Pharmaceutics · 2026