Journal of Indian Prosthodontic Society

Evaluation and comparison of the scanning accuracy of three intraoral scanners for single crown, three-unit, and cross-arch fixed partial denture by three-dimensional digital superimposition technique and measurement of vertical marginal fit of three-dimensional-printed resin retainers: An in vitro study.

Mansi Dhiran, Ajay Sabane, Abhijit Ashok Patil, Arti Gachake, Darshana Mundhe, Rupali Patil

Department of Prosthodontics and Crown and Bridge, Bharati Vidyapeeth Deemed to be University Dental College and Hospital, Pune, Maharashtra, India.

Preclinical

Key points

From the paper's conclusion
  • All three scanners delivered clinically acceptable results; 3Shape TRIOS 3 yielded the highest accuracy and best marginal adaptation. Accuracy and marginal fit deteriorated as prosthesis span increased.

Abstract

AIM: This study aims to evaluate and compare the scanning accuracy of three IOSs and measuring the vertical marginal fit of three-dimensional (3D) printed resin retainers for single crown, three-unit, and cross-arch fixed partial denture situations. SETTINGS AND DESIGN: Comparative In Vitro Study. MATERIALS AND METHODS: A maxillary Nissin typodont with prepared abutments at #13, #23, #25, #27, and #16 served as the master model. A calibrated Shining 3D laboratory scanner provided the reference standard tessellation language (control). Each IOS generated six scans ( n = 6/scanner; 18 test scans). Trueness and precision were derived by best-fit superimposition in Geomagic Design X. Fifty-four single-unit, three-unit, and cross-arch retainers were 3D-printed. Vertical marginal discrepancy was measured at mid-buccal and mid-lingual sites using a digital stereomicroscope (×20). STATISTICAL ANALYSIS USED: One-way ANOVA with Tukey's post-hoc test was used for comparisons among scanners and prosthesis spans. Pearson's correlation was used to assess the relationship between trueness and marginal fit, and Chi-square test was used to assess clinical acceptability. A P value ≤0.05 was considered statistically significant. RESULTS: 3Shape TRIOS 3 showed the highest trueness (28.5 ± 5.2 µm) and precision (18.2 ± 3.1 µm), followed by Medit i700 and Shining 3D Aoralscan 3 ( P < 0.001). Marginal discrepancy was lowest with 3Shape TRIOS 3 (52.4 ± 6.5 µm) and increased with span. A positive correlation was found between trueness and marginal fit ( r = 0.82; P < 0.001). Clinical acceptability was 100%, 94.4%, and 77.8% for 3Shape TRIOS 3, Medit i700, and Shining 3D Aoralscan 3, respectively ( P = 0.041). CONCLUSION: All three scanners delivered clinically acceptable results; 3Shape TRIOS 3 yielded the highest accuracy and best marginal adaptation. Accuracy and marginal fit deteriorated as prosthesis span increased.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Evaluation and comparison of the scanning accuracy of three intraoral scanners for single crown, three-unit, and cross-arch fixed partial denture by three-dimensional digital superimposition technique and measurement of vertical marginal fit of three-dimensional-printed resin retainers: An in vitro study — Science Skim