Prospective multicentric validation of a novel uroflowmetry device (Pflow™) against conventional gravimetric uroflowmetry.
Arvind P Ganpule, Ashish Rawandale-Patil, Kavan Anil Takvani, Lokesh G Patni, Preeti Ashish Rawandale, Tanisha Pranish Kushare
Department of Urology, Muljibhai Patel Urological Hospital, Nadiad, Gujarat, India. · Department of Urology, Institute of Urology, Dhule, Maharashtra, India.
Key points
From the paper's conclusion- Pflow™ demonstrates good agreement with conventional uroflowmetry and may serve as a simple adjunct for monitoring urinary flow in clinical settings.
Abstract
BACKGROUND: Conventional gravimetric uroflowmetry, although reliable, is restricted to clinical settings and may be influenced by inhibition and logistical constraints. The present study device is a gravity-driven volumetric uroflowmetry device that uses a calibrated multichamber collection system to derive standard uroflowmetry parameters without requiring electronics or external power. This study was done to assess the agreement between a novel portable uroflowmetry device (Pflow™) and conventional gravimetric uroflowmetry. MATERIALS AND METHODS: This prospective two-center, comparative study was conducted over 12 months (July 2023-July 2024) across two tertiary urology centers and included 372 patients (292 males; mean age 50.3 ± 18.5 years). Measurements were recorded simultaneously using both the devices during a single voiding event. Parameters analyzed included voided volume, peak flow rate (Qmax), average flow rate (Qavg), and voiding time. Agreement was assessed using the Wilcoxon signed-rank test, Spearman correlation, intraclass correlation coefficient (ICC), and Bland-Altman analysis. RESULTS: No statistically significant differences were observed between the two methods for voided volume (271 mL vs. 270 mL, P = 0.440), voiding time (40 s vs. 38 s, P = 0.204), Qmax (15 mL/s vs. 14.4 mL/s, P = 0.209), and Qavg (9 mL/s vs. 10 mL/s, P = 0.739) and strong correlations were observed (ρ = 0.799-0.963, P < 0.001). ICC demonstrated excellent agreement for voided volume (0.963) and good agreement for all measured flow parameters. Bland-Altman analysis showed minimal bias. CONCLUSION: Pflow™ demonstrates good agreement with conventional uroflowmetry and may serve as a simple adjunct for monitoring urinary flow in clinical settings.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Efficacy and Reliability of Mobile Uroflowmetry in Patients With Benign Prostatic Hyperplasia Undergoing Transurethral Resection: Prospective Multicenter Observational Pilot Validation StudyJournal of Medical Internet Research · 2025
- Automated monitoring of urine output in hospitalized patients with indwelling urinary catheters: a clinical evaluation on the cardiology wardNetherlands Heart Journal · 2026
- Impact of a Pre-Visit Reminder Program on Outpatient Uroflowmetry Test Validity: A Retrospective Chart Review.Urology · 2026
- Standing or Sitting to Void? Comparative Uroflowmetry in Healthy Male Volunteers: A Randomized Crossover Study with Equivalence Analysis.Neurourology and urodynamics · 2026
- Closing the documentation gap: accuracy and completeness of a novel EHR-integrated gravitational urine production gauge vs. manual chartingEuropean Heart Journal · 2026
- Modern Diagnostic Tools for Benign Prostatic Hyperplasia.Current urology reports · 2026