Severe Rotisserie Tilt of Scleral-Fixated CT Lucia 602 Intraocular Lenses: A Multicenter Case Series.
Dipen Kumar, Jeffrey Tan, Jacqueline G Jeon-Chapman, Kamran M Riaz, Yingna Liu, Chirag Shah, Brenna Bullock, Mark K Walsh,
Department of Ophthalmology, The University of North Carolina at Chapel Hill, Chapel Hill, North Carolina. · Retina-Vitreous Associates Medical Group, Los Angeles.
Key points
From the paper's conclusion- This multicenter case series describes severe rotisserie tilt of scleral-fixed CT Lucia 602 intraocular lenses (IOL) observed across multiple surgeons and centers between 2021 to 2023.
Abstract
This multicenter case series describes severe rotisserie tilt of scleral-fixed CT Lucia 602 intraocular lenses (IOL) observed across multiple surgeons and centers between 2021 to 2023. The study reviewed 1,408 intrascleral haptic fixation (ISHF) surgeries performed by 20 surgeons from 18 centers, including both retina and anterior segment specialists, at both academic and private practice settings, and identified 45 tilted IOL cases with the CT Lucia requiring surgical intervention. There was a wide variation in instrumentation, tunnel construction approaches, and lens insertion techniques. The average estimated tilt was 74.8º ± 40.1º. Abnormalities in the haptic-optic junction were found in 34 of 45 (76%) cases. Management included replacing the IOL, placing an anterior chamber IOL, refixating the IOL, or leaving the patient aphakic. Of those receiving another CT Lucia lens, seven of 16 (44%) cases experienced repeat tilt, representing a meaningful risk of recurrence. Given the heterogeneity of surgeon technique, our findings may represent a concerning reproducible failure phenotype.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Late Reposition of a Dislocated Toric Intraocular Lens in a Fibrosed Capsular Bag: A Case Report.Case reports in ophthalmology · 2026
- Haptic-Extended Flange Intrascleral Fixation for Single-Piece Acrylic Intraocular Lenses.Journal of clinical medicine · 2026
- Outcomes of Carlevale-type scleral-fixated intraocular lenses as a stable platform in silicone oil-filled eyes: a multicenter studyScientific Reports · 2026
- Early postoperative rotational behavior of a plate-haptic multifocal intraocular lensBMC Ophthalmology · 2026
- Comparative Two-Year Outcomes of Three Different Three-Piece Intraocular Lenses in Flanged Intrascleral Fixation.Medicina (Kaunas, Lithuania) · 2026
- Recurrent Iris Capture after Iris Suturing following Scleral-Sutured Intraocular Lens Fixation: A Case Report.Case reports in ophthalmology · 2026