Surgically Induced Astigmatism After PRESERFLO MicroShunt Ab-Externo Implantation in Microinvasive Bleb Surgery.
Colya N Englisch, Karl T Boden, Peter Szurman, Charlotte Semoulin, André M Trouvain, Anna Theresa Fröhlich, Coline Englisch, André Messias,
Eye Clinic Sulzbach, Knappschaft Hospitals Saar, Sulzbach, Germany. · Department of Experimental Ophthalmology, Saarland University, Homburg, Germany.
Key points
From the paper's conclusion- PRESERFLO MicroShunt implantation induces a small but significant refractive astigmatism vector directed toward the vertical meridian, sustained over 12 months. PSA changes demonstrated a measurable vectorial shift but were small in magnitude, ultimately appearing unlikely to fully explain the refractive SIA, suggesting that non-corneal mechanisms related to intraocular pressure reduction might be critical contributors.
Abstract
PURPOSE: To investigate surgically induced astigmatism (SIA) after stand-alone PRESERFLO MicroShunt (Santen Pharmaceutical Co Ltd) ab-externo implantation. METHODS: This retrospective study included eyes that had mandatory preoperative and 3-month subjective refraction available, with 12-month data collected when available. Tomography measurements, when available, were obtained using Scheimpflug imaging. Refractive and anterior (ASA) and posterior (PSA) surface astigmatism were evaluated using vector analysis. RESULTS: A total of 117 eyes from 108 patients were included; 12-month data were available for 57 eyes. Refractive cylinder showed an increase from 1.37 ± 0.91 diopters (D) at baseline to 1.62 ± 1.06 D at 3 months (P = .01), which was sustained at 12 months (P > .99). From baseline to 3 months, the centroid SIA was 0.47 D at 78.00°. When restricting to the 57 eyes with 12-month follow-up data, this gave a centroid SIA of 0.61 D at 76.22° from baseline to 3 months and 0.38 D at 82.57° from 3 to 12 months. In 25 eyes with preoperative and 3-month tomography data, centroid SIA was 0.17 D at 43.06° for the ASA and 0.35 D at 92.29° for the PSA. CONCLUSIONS: PRESERFLO MicroShunt implantation induces a small but significant refractive astigmatism vector directed toward the vertical meridian, sustained over 12 months. PSA changes demonstrated a measurable vectorial shift but were small in magnitude, ultimately appearing unlikely to fully explain the refractive SIA, suggesting that non-corneal mechanisms related to intraocular pressure reduction might be critical contributors.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Refractive and Corneal Astigmatism After Implantation of a Supraciliary Drainage Device in Microinvasive Glaucoma SurgeryJournal of Refractive Surgery · 2026
- Comparison of Surgically Induced Astigmatism After Glaucoma Filtering and Implant Procedures.Clinical ophthalmology (Auckland, N.Z.) · 2026
- Surgically induced astigmatism after phacotrabeculectomy: a prospective observational study at ECWA Eye Hospital, Kano.BMC ophthalmology · 2026
- Corneal versus refractive surgically induced astigmatism: vector analysis of their dissociation and implications for toric IOL planning.American journal of ophthalmology · 2026
- Comparison of one-year postoperative outcomes between PreserFlo MicroShunt and trabeculectomy in Japanese patients with open angle glaucoma: a retrospective cohort study using propensity score matchingJapanese Journal of Ophthalmology · 2026
- Association between distance from the angle recess to the tube entry site and corneal endothelial cell loss after PreserFlo MicroShunt implantation.Japanese journal of ophthalmology · 2026