Viscoelastic-Assisted Patient Interface Docking: A Technical Optimization in LenSx Femtosecond Laser-Assisted Cataract Surgery.

Danmin Cao, Wenjing Luo, Juan Li, Tingting Peng, Huafang Guo, Anqi Xie, Ying Sun, Yong Wang

Department of Ophthalmology, The First Affiliated Hospital of Guangzhou Medical University. · Aier Eye Hospital of Wuhan University, Wuhan, Hubei Province.

Randomised · n=100

Key points

From the paper's conclusion
  • Viscoelastic-assisted patient interface docking in FLACS effectively elevates one-time docking success rate, reduces docking attempts, and shortens docking time.

Abstract

PURPOSE: To evaluate the efficacy of viscoelastic-assisted patient interface docking in LenSx (Alcon Laboratories, Inc) femtosecond laser-assisted cataract surgery (FLACS). METHODS: This was a randomized controlled trial. Patients undergoing FLACS from January to August 2025 at Aier Eye Hospital of Wuhan University were randomized via a random number table to receive balanced salt solution (BSS) or visoeleastic as the patient interface docking medium. The primary outcome was docking efficiency, measured by one-time docking success rate, the number of docking attempts, and mean docking time. Secondary outcomes included surgical safety (subconjunctival hemorrhage, capsulotomy completeness/tear rate), laser treatment duration (anterior capsulotomy time, nucleus pretreatment time, total laser emission time), and patient comfort (post-laser pain sensation). RESULTS: A total of 100 patients were enrolled, 50 in each group. Suction loss occurred in 7 patients (14%) in the BSS group and 1 patient (2%) in the viscoelastic group; the one-time docking success rate was significantly higher in the viscoelastic group (98%) than in the BSS group (86%) (chi-square = 3.93, P < .05). The viscoelastic group also had fewer mean docking attempts (1.02 ± 0.14) than the BSS group (1.16 ± 0.42), showing a significant difference (t = 2.23, P < .05). The viscoelastic group exhibited significantly shorter mean docking time (44.66 ± 4.47 seconds) compared to the BSS group (48.62 ± 3.11 seconds) (t = 2.17, P < .05). No significant differences were observed between the groups in subconjunctival hemorrhage, capsulotomy completeness/tear rate, anterior capsulotomy time, nucleus pretreatment time, total femtosecond laser emission time, or patient-reported pain sensation (all P > .05). CONCLUSIONS: Viscoelastic-assisted patient interface docking in FLACS effectively elevates one-time docking success rate, reduces docking attempts, and shortens docking time.

Read the paper