Viscoelastic-Assisted Patient Interface Docking: A Technical Optimization in LenSx Femtosecond Laser-Assisted Cataract Surgery.
Danmin Cao, Wenjing Luo, Juan Li, Tingting Peng, Huafang Guo, Anqi Xie, Ying Sun, Yong Wang
Department of Ophthalmology, The First Affiliated Hospital of Guangzhou Medical University. · Aier Eye Hospital of Wuhan University, Wuhan, Hubei Province.
Key points
From the paper's conclusion- Viscoelastic-assisted patient interface docking in FLACS effectively elevates one-time docking success rate, reduces docking attempts, and shortens docking time.
Abstract
PURPOSE: To evaluate the efficacy of viscoelastic-assisted patient interface docking in LenSx (Alcon Laboratories, Inc) femtosecond laser-assisted cataract surgery (FLACS). METHODS: This was a randomized controlled trial. Patients undergoing FLACS from January to August 2025 at Aier Eye Hospital of Wuhan University were randomized via a random number table to receive balanced salt solution (BSS) or visoeleastic as the patient interface docking medium. The primary outcome was docking efficiency, measured by one-time docking success rate, the number of docking attempts, and mean docking time. Secondary outcomes included surgical safety (subconjunctival hemorrhage, capsulotomy completeness/tear rate), laser treatment duration (anterior capsulotomy time, nucleus pretreatment time, total laser emission time), and patient comfort (post-laser pain sensation). RESULTS: A total of 100 patients were enrolled, 50 in each group. Suction loss occurred in 7 patients (14%) in the BSS group and 1 patient (2%) in the viscoelastic group; the one-time docking success rate was significantly higher in the viscoelastic group (98%) than in the BSS group (86%) (chi-square = 3.93, P < .05). The viscoelastic group also had fewer mean docking attempts (1.02 ± 0.14) than the BSS group (1.16 ± 0.42), showing a significant difference (t = 2.23, P < .05). The viscoelastic group exhibited significantly shorter mean docking time (44.66 ± 4.47 seconds) compared to the BSS group (48.62 ± 3.11 seconds) (t = 2.17, P < .05). No significant differences were observed between the groups in subconjunctival hemorrhage, capsulotomy completeness/tear rate, anterior capsulotomy time, nucleus pretreatment time, total femtosecond laser emission time, or patient-reported pain sensation (all P > .05). CONCLUSIONS: Viscoelastic-assisted patient interface docking in FLACS effectively elevates one-time docking success rate, reduces docking attempts, and shortens docking time.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Femtosecond laser-assisted cataract surgery versus phacoemulsification in shallow anterior chambers: a systematic review and meta-analysis.Journal of cataract and refractive surgery · 2026
- Comparing Outcomes of Femtosecond Laser-Assisted Cataract Surgery and Conventional Phacoemulsification in a Randomized Controlled Trial.Journal of visualized experiments : JoVE · 2026
- Effect of temperature variation on the corneal endothelial cell during femtosecond laser-assisted cataract surgery compared with conventional phacoemulsification cataract surgery: A prospective studyMedicine · 2026
- Centration and Torsion Control Following Lenticule Creation With Keratorefractive Lenticule Extraction: Effect on Aberration and Astigmatism.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026
- Flap Optimization Using VISUMAX 800 Femtosecond Laser for Laser-Assisted In Situ Keratomileusis.Journal of clinical medicine · 2026
- Incision-induced astigmatic changes in eyes undergoing simultaneous phakic intraocular lens explantation and phacoemulsificationJournal of Cataract & Refractive Surgery · 2026