Thermal Skin Burns Following Infrared Vein Finder Application in Pediatric Patients: A Report of 2 Cases and Protocol Revision.
Sujata S Rawlani, Nandini Banad, Jenil Sheth, Raja Narsingrao
Shantilal Shanghvi Eye Institute, Mumbai, Maharashtra, India. · LV Prasad Eye Institute, Hyderabad, Telangana, India.
Key points
From the paper's conclusion- These cases highlight a patient safety concern regarding pocket flashlight-based vein finders. We advocate for institutional protocol revisions including mandatory time limits for device application, preoperative skin assessments, and preferential use of LED-based devices with established safety profiles. Awareness of device-specific thermal risks is essential to prevent iatrogenic harm.
Abstract
BACKGROUND: Peripheral venous cannulation remains a cornerstone of pediatric anesthesia practice. Infrared vein finder devices have gained popularity as aids to improve venous visualization in children with difficult access. However, the thermal hazards associated with certain device types, particularly pocket flashlight-based models, remain underappreciated among clinical staff. CASE PRESENTATION: We report 2 pediatric patients who sustained grade 2 thermal burns at peripheral venous access sites following the use of a pocket flashlight-based infrared vein finder during ophthalmic surgical procedures. The first patient, an 8-month-old girl born at term with appropriate developmental milestones undergoing congenital cataract surgery, developed a ruptured blister on her left elbow after 45 minutes of intermittent device application across multiple sites. The second patient, a 1-year-old boy, born preterm at 32 weeks gestation (birth weight of 1.9 kg) with delayed milestones scheduled for examination under anesthesia for congenital glaucoma, developed thermal blisters on his right palm. Retrospective review of operating room closed-circuit television footage confirmed the device had been in continuous contact with the right palm for approximately 5 minutes during the initial cannulation attempt. Notably, the contralateral palm remained unaffected, where device contact lasted only 2 minutes. Both injuries occurred in the absence of any underlying dermatological conditions. CONCLUSION: These cases highlight a patient safety concern regarding pocket flashlight-based vein finders. We advocate for institutional protocol revisions including mandatory time limits for device application, preoperative skin assessments, and preferential use of LED-based devices with established safety profiles. Awareness of device-specific thermal risks is essential to prevent iatrogenic harm.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- What barriers and facilitators influence the implementation of infrared vein visualization technology for pediatric peripheral intravenous catheterization? A TDF-informed qualitative study.Frontiers in pediatrics · 2026
- Infrared Technology for Vein Visualization: Device Development.Bioengineering (Basel, Switzerland) · 2026
- Effectiveness of Transillumination Device on Reducing Peripheral Intravenous Catheter Placement Complications in Pediatric Patients: A Randomized Controlled Trial.The journal of nursing research : JNR · 2026
- Adjunctive Imaging to Confirm Ultrasound-Guided Paravertebral Catheter Tip Position in Children: Case Reports.A&A practice · 2026
- Immersive Virtual Reality for Pediatric Procedural Pain and Anxiety: A Structured Narrative Review and Developmentally Informed Framework for Clinical Matching and AI-Supported Personalization.Nursing reports (Pavia, Italy) · 2026
- Toward Real-Time Detection of Intravenous Infiltration Using a Low-Cost Strain-Sensing Patch: Validation in Animal Models and Pediatric Clinical Monitoring.IEEE journal of translational engineering in health and medicine · 2026