From the paper's conclusion- A strain-sensing patch integrated with a real-time-capable, event-based detection algorithm enabled early IV infiltration detection at low infused volumes while maintaining a false-positive alert burden that is acceptable for early-stage clinical evaluation.
Abstract
OBJECTIVE: To develop and evaluate a low-cost, skin-safe strain-sensing patch and real-time-capable algorithm for early detection of intravenous (IV) infiltration during continuous monitoring. METHODS: A custom strain sensor composed of discrete piezoresistive elements was encased between two transparent medical dressings and mounted adjacent to the IV site. Data were collected during controlled IV infiltrations in juvenile pigs and during inpatient monitoring in pediatric patients without infiltrations. A strictly causal, event-based detection algorithm was implemented using features derived from the strain signal and its first difference. Performance was evaluated using event sensitivity, time-to-detection (TTD), and false-positive alerts per hour (FP/hr). Candidate thresholds were derived from the animal training-set evidence distribution. The operating point was selected using training-data sensitivity, detection time, and control false-alert burden, then fixed before held-out testing. RESULTS: Across 16 animal recordings comprising 35 infiltration events, the algorithm detected 30 events (event sensitivity = 0.86) with a median TTD of 57 seconds at the selected operating threshold. Approximately 86.7% of detected events occurred within two minutes of infiltration onset. On the held-out recording subset, 8 of 9 infiltration events were detected, with a median TTD of 53.5 seconds. Evaluation on 117.8 hours of pediatric inpatient control data demonstrated a threshold-dependent false-positive alert burden, operating at approximately 1.2 alerts per hour at the selected operating threshold. CONCLUSION: A strain-sensing patch integrated with a real-time-capable, event-based detection algorithm enabled early IV infiltration detection at low infused volumes while maintaining a false-positive alert burden that is acceptable for early-stage clinical evaluation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Statistical analysis plan for the Preventing InfusAte injuries Throughout a Childs Hospitalisation (PATCH) trial: a type 1 hybrid randomised controlled trialmedRxiv · 2026
- Peripheral intravenous waveform analysis - a novel tool to confirm intravascular catheter placement.Journal of clinical monitoring and computing · 2026
- Clinician assessment of an in situ peripheral intravenous catheter: using ultrasound to prevent unrecognized extravasation.British journal of nursing (Mark Allen Publishing) · 2026
- Factors associated with extravasation severity in Thai patients receiving noncytotoxic intravenous infusions: An observational study.Belitung nursing journal · 2026
- Feasibility Study of Noninvasive Subcutaneous Imaging for Vein LocalizationElectronics · 2026
- Development of a method for vancomycin monitoring in interstitial fluid using a novel porous microneedle array patch.Drug metabolism and pharmacokinetics · 2026