Ethoxyquin - Determination of 2,2,4-trimethyl-6(2H)-quinolinone in urine by UPLC-MS/MS: Biomonitoring Method - Translation of the German version from 2026.
Gerhard Scherer, Nikola Pluym, Max Scherer, Nadine Rögner, Markus Stoeckelhuber, Thomas Jäger, Thomas Göen, Andrea Hartwig,
ABF Analytisch-biologisches Forschungslabor GmbH Semmelweisstraße 5 82152 Planegg Germany. · BASF SE, Corporate Health Management Carl-Bosch-Straße 38 67056 Ludwigshafen Germany.
Key points
From the paper's conclusion- The working group "Analyses in Biological Materials" of the German Senate Commission for the Investigation of Health Hazards of Chemical Compounds in the Work Area (MAK Commission) developed and verified this biomonitoring method for the determination of the most important urinary metabolite of ethoxyquin - 2,2,4‑trimethyl‑6(2H)-quinolinone (EQI).
Abstract
The working group "Analyses in Biological Materials" of the German Senate Commission for the Investigation of Health Hazards of Chemical Compounds in the Work Area (MAK Commission) developed and verified this biomonitoring method for the determination of the most important urinary metabolite of ethoxyquin - 2,2,4‑trimethyl‑6(2H)-quinolinone (EQI). Ethoxyquin is a quinoline-based synthetic antioxidant. Its use as active substance in pesticides was prohibited in 2011. Nevertheless, it continues to be used as a feed additive and is mainly introduced into the environment by feeding treated fishmeal to fish in aquaculture. A human metabolism study showed that ethoxyquin is first metabolised to 1,2‑dihydro-2,2,4‑trimethyl-6‑quinolinol (hydroxyquin) and is then further oxidised to the more stable EQI. In the biomonitoring method presented here, the glucuronide of hydroxyquin in the urine sample is enzymatically hydrolysed using β‑glucuronidase from E. coli. Thereafter, hydroxyquin oxidises spontaneously to EQI. The hydrolysate is subject to salt-assisted liquid-liquid extraction (SALLE) with ethyl acetate. Analysis is performed by UPLC‑MS/MS after positive electrospray ionisation (ESI+). Calibration is performed using standards prepared in pooled urine and processed analogously to the samples. EQI‑D10 is applied as an internal standard. The method provides reliable and accurate analytical results, as shown by the good precision data with standard deviations no greater than 6%. Good accuracy data were obtained with mean relative recoveries in the range of 103-110%. The method is both selective and sensitive, whereby a quantitation limit of 0.03 μg EQI/l was achieved.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Tebuconazole and penconazole - Determination of TEB-OH, TEB-COOH, PEN-OH, and PEN-COOH in urine by LC-MS/MS: Biomonitoring Method - Translation of the German version from 2026.The MAK collection for occupational health and safety · 2026
- Biomonitoring: a demonstration of the importance of direct analytical methods for the assessment of chemical exposures.Frontiers in chemistry · 2026
- Determination of Quizalofop (DPX-79376) Residues in Hemp Cannabis Matrices by LC-MS/MSACS Agricultural Science & Technology · 2026
- Comprehensive chemical characterization of quanduzhong capsules and In Vivo profiling of parent constituents and metabolites in rats by UPLC-Q-TOF-MSE.Journal of pharmaceutical and biomedical analysis · 2026
- Simultaneous determination of the proinsecticide flometoquin and its bioactive metabolite in plant-derived foodstuffs by modified QuEChERS coupled with UHPLC-MS/MS.Journal of chromatography. A · 2026
- Phase Behavior and Separation Performance of Ethyl Lactate/Carbohydrate or Salt ATPS for Cypermethrin, Amoxicillin, and 17β-EstradiolACS Omega · 2026