Anatomical Risk Factors for Technical Failure of the Perclose Vascular Closure Device: A Single-Center Retrospective Study.
Hirofumi Jinno, Shinsuke Kikuchi, Tsutomu Doita, Izumi Fukii, Kyokei Fuchizawa, Takayuki Uramoto, Seima Ohira, Keisuke Kamada,
Department of Vascular Surgery, Asahikawa Medical University, Asahikawa, Hokkaido, Japan. · Department of Cardiovascular Surgery, Osaka University, Suita, Osaka, Japan.
Key points
From the paper's conclusion- Technical failure was driven by unfavorable CFA anatomy-heavy calcification, a small diameter, and a shallow puncture depth-and by emergency procedures. Preprocedural imaging assessment is essential; alternative closure or surgical cutdown should be considered in unfavorable or emergent cases.
Abstract
OBJECTIVES: Percutaneous large-bore femoral closure with the Perclose device is widely used, but technical failure remains clinically important. We identified anatomical risk factors for technical failure. METHODS: We retrospectively analyzed endovascular aortic repair and endovascular therapy procedures using the Perclose device at the common femoral artery (CFA) between April 2020 and March 2025. Preoperative computed tomography (CT) measured CFA diameter, puncture depth, and a calcification score (0-4). Technical failure was defined as failure to achieve hemostasis with the intended deployment. Independent predictors were identified using patient-clustered generalized estimating equations. RESULTS: Among 300 patients (485 limbs), technical failure occurred in 52 limbs (10.7%), of which 75.0% required surgical repair. Independent predictors were a higher calcification score (odds ratio [OR] 2.13 per 1-point increase; 95% confidence interval [CI] 1.37-3.31), an emergency procedure (OR 3.77; 1.55-9.14), a smaller CFA diameter (OR 1.16 per 1-mm decrease; 1.01-1.32), and a shallower puncture depth (OR 1.22 per 5-mm decrease; 1.01-1.47); the area under the curve was 0.72. CONCLUSIONS: Technical failure was driven by unfavorable CFA anatomy-heavy calcification, a small diameter, and a shallow puncture depth-and by emergency procedures. Preprocedural imaging assessment is essential; alternative closure or surgical cutdown should be considered in unfavorable or emergent cases.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Sheath-to-Artery Diameter Ratio: A Robust Predictor of Early Access-Site Adverse Events in Percutaneous Procedures.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- Clinical outcomes and cost analysis of plug-based versus suture-based vascular closure devices for percutaneous large bore aortic proceduresVASA · 2026
- Comparison of the Technical Success of Obtura and StarClose Systems for Femoral Arterial Access-Site Closure in Neurointerventional Procedures: A Single-Center Retrospective Cohort Study.Journal of clinical medicine · 2026
- Identifying a Favorable Morphological Profile for Transcatheter Atrial Septal Defect Closure in Patients With Deficient Posteroinferior Rims.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- Femoral Artery Access Outcomes and Technical Considerations in Vascular Ehlers-Danlos Syndrome.Journal of vascular surgery · 2026
- Plug-based versus suture-based vascular closure devices in large-bore femoral access for endovascular repair: a systematic review and meta-analysisFrontiers in Cardiovascular Medicine · 2026