Ex Vivo Evaluation of Endothelial and Adventitial Temperature during Endovenous Laser Ablation Using a 1470-nm Radial 2ring Slim Fiber: Porcine Jugular Vein Model.
Osamu Suzuki
Suzuki Medical Clinic, Suzuka, Mie, Japan.
Key points
From the paper's conclusion- In this ex vivo model, 6.5W with a LEED of 65 J/cm tended to produce more sustained endothelial and adventitial thermal exposure, whereas 5W with a LEED of 25 J/cm tended to produce lower adventitial exposure. Although further clinical validation is required, segment-specific settings may be reasonable.
Abstract
OBJECTIVES: Endovenous laser ablation (EVLA) using a 1470-nm radial 2-ring slim fiber is widely performed, but optimal ablation settings remain empirical. This ex vivo study aimed to evaluate the effects of ablation conditions on endothelial and adventitial temperature dynamics, providing insights into both treatment efficacy and safety. METHODS: Porcine jugular veins were used in an ex vivo model. EVLA was performed using a 1470-nm diode laser with the 2-ring slim fiber under 2 conditions: 5W with a linear endovenous energy density (LEED) of 25 J/cm and 6.5W with a LEED of 65 J/cm. Endothelial and adventitial temperatures were continuously measured at 0.1-second intervals. Cumulative durations with endothelial temperature ≥70°C and adventitial temperature ≥43°C were assessed. RESULTS: Endothelial temperatures reached ≥70°C under both conditions, with a longer duration observed at 6.5W. Cumulative adventitial exposure ≥43°C was prolonged at 6.5W, whereas it remained minimal at 5W. CONCLUSIONS: In this ex vivo model, 6.5W with a LEED of 65 J/cm tended to produce more sustained endothelial and adventitial thermal exposure, whereas 5W with a LEED of 25 J/cm tended to produce lower adventitial exposure. Although further clinical validation is required, segment-specific settings may be reasonable.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Assessment of temperature dynamics in pulsed field ablation with a variable-loop circular catheter: a comparative analysis of waveform configurations and irrigation rates in specimens of bovine ventricular myocardiumEP Europace · 2025
- The impact of "flush" endovenous laser ablation on thromboembolic risk: a retrospective safety analysis.Phlebology · 2026
- Pulsed Field Ablation Therapy Parameters: Enhancing Lesion Depth While Reducing Musculoskeletal Stimulation and Thermal Response in an Irrigated Tip Catheter: A Comprehensive Study Using In Silico, In Vitro, and In Vivo Models.Circulation. Arrhythmia and electrophysiology · 2026
- Endovenous thermal ablation combined with foam sclerotherapy is a safe and effective method for the treatment of lower extremity varicose veins: a multicenter retrospective analysis.Scientific reports · 2026
- Endovenous laser ablation versus radiofrequency ablation in chronic venous insufficiency: a comparative meta-analysis of efficacy and safety.Frontiers in medicine · 2026
- Testbed Calibration for Developing Wireless Power Solutions in Heart Failure LVAD Treatment.Sensors (Basel, Switzerland) · 2026