Testbed Calibration for Developing Wireless Power Solutions in Heart Failure LVAD Treatment.
Omar J Escalona, Matthew Hutchinson, David McEneaney, Mohammad L Karim
School of Engineering, Ulster University, Belfast BT15 1AP, UK. · BDIC MedTech Ltd., School of Engineering, Ulster University, Belfast BT15 1AP, UK.
Key points
From the paper's conclusion- This study investigates a practical method for calibrating a thermostatically controlled, actively circulated isotonic saline (0.9% w/v) in a testbed system environment against the in vivo thermal profiles previously reported in porcine models for a two-channel 200 kHz pulsed RF TETS.
Abstract
A transcutaneous energy transfer system (TETS) offers a route to eventually eliminating the need for a percutaneous driveline, the principal source of infection in left ventricular assist device (LVAD) therapy, but localised tissue heating still limits clinical translation. ISO 14708-1 restricts the local tissue temperature rise in an active implantable device to ΔT ≤ 2 °C, a margin that high-power pulsed radio frequency (RF) transmission approaches at clinically useful, implanted Li-ion battery charging rates. Establishing whether a given pulsed transmission protocol is thermally safe has so far depended on in vivo measurements. This study investigates a practical method for calibrating a thermostatically controlled, actively circulated isotonic saline (0.9% w/v) in a testbed system environment against the in vivo thermal profiles previously reported in porcine models for a two-channel 200 kHz pulsed RF TETS. The calibration process follows a comprehensive three-phase approach encompassing dry baseline characterisation, wet calibration, and extended 4 h validation at a calibrated saline flow rate of 1.667 L/min, selected from three pump settings. Under the 50 V/320 ms/5 s protocol, the testbed reproduced the porcine sawtooth morphology, a peak secondary coil ΔT of 2.824 ± 0.20 °C and a charging rate of 2494 J/h (±16%, k = 2), comparable to, though not demonstrated equivalent to, the in vivo benchmark at equal transmitter voltage. Its mean plateau ΔT was 0.7-1.2 °C lower, and its post-cessation relaxation four to six times faster than in vivo, indicating that the testbed removes heat more effectively than perfused tissue; fluid flow rate is proposed as the calibration parameter controlling this bias, pending a study with continuously adjustable flow. A 4 h session at 36 °C showed no heat accumulation and kept the array-mean secondary ΔT below 2 °C; however, the thermistor placed next to the secondary coil centre reached 2.27 °C, thus, in this aspect, the protocol did not meet the ISO 14708-1 local limit.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Assessment of temperature dynamics in pulsed field ablation with a variable-loop circular catheter: a comparative analysis of waveform configurations and irrigation rates in specimens of bovine ventricular myocardiumEP Europace · 2025
- An Open-Source Dual-Channel RTD Thermal Control Platform for Research-Scale PEM Water Electrolysis Balance-of-Plant IntegrationEnergies · 2026
- Fluid Antennas for Non-Monotonic THz-WPT With Channel Estimation ErrorsIEEE Journal of Selected Topics in Signal Processing · 2026
- Assumption-bounded reduced electro-thermal modeling of lesion formation in radiofrequency cardiac ablation: power-time allocation, thermal latency, perfusion, and temperature limitation.Computer methods in biomechanics and biomedical engineering · 2026
- A Flexible Thermoelectric Generator with Synergistic Thermal Management and Adaptive Power Conditioning for Autonomous Healthcare Monitoring.ACS applied materials & interfaces · 2026
- StimulHeat: a low-energy wearable thermal feedback device using Peltier elements with heat flow controlled loop for hand interactions in virtual reality.HardwareX · 2026