Within-Session Repeatability of Spectral-Domain OCT Angiography and Inter-Device Agreement with Swept-Source OCT Angiography in Diabetic Eyes
Ruibo Chen, Jian Xu, Shanhu Qiu, Zilin Sun, Jie Luan, Zhonghong Zhang, Yuzhi Ding
Key points
From the paper's conclusion- Purpose: The aim of this study was to assess the repeatability of a spectral-domain (SD) OCTA device, the reproducibility of its measurements between raters, and its agreement with a swept-source (SS) OCTA device in diabetic eyes.
Abstract
Purpose: The aim of this study was to assess the repeatability of a spectral-domain (SD) OCTA device, the reproducibility of its measurements between raters, and its agreement with a swept-source (SS) OCTA device in diabetic eyes. Methods: Each right eye underwent three consecutive 12 × 12 mm macular OCTA scans on the SD-OCTA device and one on the SS-OCTA device, acquired in the same session by the same examiner. Retinal thickness and vascular density in the fovea, parafovea and perifovea, and the areas of the foveal avascular zone (FAZ) and capillary non-perfusion (CNP), were exported for analysis. Repeatability and inter-rater reproducibility were assessed by ICCs, and inter-device agreement by Bland–Altman analysis. Results: A total of 282 right eyes of 282 participants with type 2 diabetes were imaged. Inter-rater reproducibility was good for retinal thickness in all ETDRS areas (ICCs 0.82–0.85) but poor for all vascular parameters (all ICCs < 0.3). Repeatability of the SD-OCTA device was excellent for retinal thickness, FAZ area and CNP area (ICCs > 0.95) and moderate for foveal vascular density (ICC = 0.77). CNP detection was concordant between the two devices (κ = 0.854, 95% CI 0.77–0.94), but the SS-OCTA device read 32.91 ± 58.35 μm (fovea) to 62.01 ± 28.50 μm (perifovea) thicker than the SD-OCTA device did (95% limits of agreement: −81.5 to 147.3 μm in the fovea and 6.2 to 117.9 μm in the perifovea). Conclusions: The SD-OCTA device gave repeatable measurements of macular thickness, FAZ area and CNP area, similar to those reported for SS-OCTA systems. The two devices also agreed on whether CNP was present. However, their quantitative values were not interchangeable, and the vascular parameters were poorly reproducible between raters. The device may be a more accessible alternative, but its value for screening has not been tested, and device-specific reference values should be used.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Peripapillary Retinal and Choroidal Vasculature in Patients with Diabetes of Different Durations without Clinical Diabetic RetinopathyDiabetes Metabolic Syndrome and Obesity · 2026
- Comparison of spectral-domain optical coherence tomography biometry measurements with partial coherence interferometry optical biometryBMC Ophthalmology · 2026
- Quantifying segmentation sensitivity in OCTA: Device-specific profiles across three commercial platformsPLoS ONE · 2026
- Different Patterns of Progression in Nonproliferative Diabetic Retinopathy: Findings From a 10-Year Cohort Study.Diabetes · 2026
- Macular Deep Capillary Plexus Ischemia as a Biomarker for Identifying Referable Diabetic RetinopathyInvestigative Ophthalmology & Visual Science · 2025
- SD-OCT vs FA in diabetic retinopathy: a sensitivity-specificity tradeoff in detecting intraretinal macroaneurysms.Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie · 2026