Sensors (Basel, Switzerland)

Airflow, Heat Transfer, and Relative-Humidity Conditioning in the Plantower PMS5003 Sensor and PurpleAir Monitor.

James R Ouimette

Independent Researcher, Santa Rosa, CA 95409, USA.

Key points

From the paper's conclusion
  • We measured air temperature in the particle detection zone (PDZ) of the Plantower PMS5003 (PMS), both standalone and within a PurpleAir (PA) monitor.

Abstract

Low-cost optical particle sensors are widely used for ambient PM2.5 measurements, but internal heating can raise sample temperature and reduce relative humidity (RH) before optical detection. We measured air temperature in the particle detection zone (PDZ) of the Plantower PMS5003 (PMS), both standalone and within a PurpleAir (PA) monitor. Laboratory measurements demonstrated greater PDZ warming within the PA than in the standalone PMS. A wraparound USB heater was applied to PAs to reproduce the low sample-RH conditions used by regulatory PM2.5 monitors and nephelometers. Field experiments in Santa Rosa, California, compared standalone PMS, unheated PA, and heated PA measurements. External heating substantially reduced PDZ RH and lowered reported PM2.5 at elevated ambient RH. Internal heating in a standard PA also reduced reported PM2.5 relative to a standalone PMS, consistent with partial removal of hygroscopic aerosol water before optical detection. Computational fluid-dynamics and heat-transfer simulations characterized internal airflow, residence time, and temperature distributions, providing a physical framework for interpreting the measurements.