Evaluation of a Non-Functionalized Fiber Taper-Based Mach-Zehnder Interferometer for Bulk Refractive-Index Monitoring Using Paracetamol as a Model Solute.
Leonardo Díaz Marulanda, Monserrat Alonso-Murias, Erika Rodríguez-Sevilla, Andrés Hernández-Pascual, David Monzon-Hernández, Carlos Galíndez Jamioy, Javier A Montoya
Departamento de Electrónica, Grupo de Óptica e Informática, Universidad Popular del Cesar, Balneario Hurtado, Vía a Patillal, Valledupar 200001, Cesar, Colombia. · Centro de Investigaciones en Óptica, A.C. (CIO), Loma del Bosque 115, Colonia Lomas del Campestre, León 37150, Guanajuato, Mexico.
Key points
From the paper's conclusion- Here, we experimentally demonstrate a non-functionalized fiber taper-based Mach–Zehnder interferometer (MZI) as a bulk refractive-index sensor using paracetamol solutions as a controlled model system.
Abstract
Accurate monitoring of composition-related changes in concentrated pharmaceutical streams is relevant to process control and quality assessment. Here, we experimentally demonstrate a non-functionalized fiber taper-based Mach-Zehnder interferometer (MZI) as a bulk refractive-index sensor using paracetamol solutions as a controlled model system. The sensing mechanism is based on changes in the bulk refractive index of the surrounding medium, which are correlated with paracetamol concentration in controlled aqueous samples prepared using either a paracetamol standard or commercial tablets. Wavelength tracking of the MZI transmission spectra over 0-12.5 mg/mL paracetamol, corresponding to a refractive-index range of 1.333-1.335 RIU, yielded a sensitivity of 233 pm/(mg/mL) and 1497.3 nm/RIU, with a limit of detection (LOD) of 1.08 mg/mL. An intensity-based interrogation provided an average absolute sensitivity of 5.73 × 10-6 W/(mg/mL), a practical resolution of 44.2 mg/L, and an improved LOD of 75.6 mg/L. Additionally, an MZI configuration in reflection, evaluated using reagent-grade paracetamol standard solutions, yielded a wavelength sensitivity of 318 pm/(mg/mL), while phase tracking of reflected spectra yielded a sensitivity of -97.4 mrad/(mg/mL). These results demonstrate the versatility, linearity, repeatability, temporal stability, and low-maintenance operation of the MZI for bulk refractive-index monitoring, confirming its performance under controlled model system conditions.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Smart Sensor Based on Multiple Cascaded Michelson Interferometers with a Long-Period Fiber Grating for Recognition of Paracetamol Concentrations in Solutions.Sensors (Basel, Switzerland) · 2026
- Integrated Mach-Zehnder Interferometer based on dual-microchannel capillary core fiber for simultaneous label-free BSA and temperature measurement.Biosensors & bioelectronics · 2026
- High-precision measurement method of liquid concentration based on differential laser self-mixing interferometry.Applied optics · 2026
- Reference-state transferability limits in IR140: working-solution spectroscopy, bleachable optical depth, and apparent femtosecond saturation.Spectrochimica acta. Part A, Molecular and biomolecular spectroscopy · 2026
- Bias-Aware Detection Limits, Calibration Transfer and Matched-Control Robustness Assessment in Dual-Parameter Photonic Refractive-Index and Temperature Sensing: A Coupled Interface-Mode Multilayer Case Study.Sensors (Basel, Switzerland) · 2026
- Noncontact Chemical and Morphological Profiling in Liquid via Fiber-Based Micro-FTIRarXiv · 2026