Bias-Aware Detection Limits, Calibration Transfer and Matched-Control Robustness Assessment in Dual-Parameter Photonic Refractive-Index and Temperature Sensing: A Coupled Interface-Mode Multilayer Case Study.
Agah Oktay Ertay, Muhammed Mustafa Ertay
Faculty of Engineering and Architecture, Department of Electrical and Electronics Engineering, Erzincan Binali Yildirim University, Yalnizbag Campus, Erzincan 24002, Türkiye. · Faculty of Engineering, Department of Electrical and Electronics Engineering, Duzce University, Konuralp Campus, Duzce 81620, Türkiye.
Key points
From the paper's conclusion- Two coupled interface resonances at 1517 and 1651 nm, with loaded Q of 232 and 208 and refractive-index (RI) sensitivities of 90.71 and 329.41 nm/RIU, are inverted by a bounded nonlinear calibration to 2.43×10-5 RIU and 0.155°C; the temperature channel reports the device temperature.
Abstract
Dual-parameter photonic sensors are usually reported through a nominal sensitivity and one detection limit, without stating which statistic that limit is or whether it survives transfer between devices. This computational study supplies that evaluation for one structure, a 36-layer one-dimensional multilayer read in transmission, whose Zak-phase-distinct TiO2/SiO2 photonic-crystal sections enclose a 600 nm analyte cavity and a 500 nm thermo-optic reference cavity, each carrying a 5 nm ITO/5 nm TiO2 nanolaminate insert. Two coupled interface resonances at 1517 and 1651 nm, with loaded Q of 232 and 208 and refractive-index (RI) sensitivities of 90.71 and 329.41 nm/RIU, are inverted by a bounded nonlinear calibration to 2.43×10-5 RIU and 0.155°C; the temperature channel reports the device temperature. Probability-of-detection limits at 1% false alarm and 95% detection are 4.36×10-5 RIU and 0.123°C; they are set by the calibration standards and the wavelength reference, not by the linewidth. Transferring one calibration between devices worsens them 81-fold and 203-fold; a three-point per-device correction removes 84-93% of that loss. A trivial control matched on wavelength, Q, transmission, thickness and RI sensitivity shows no topological robustness advantage. Applied to the design itself, the same evaluation shows that the modes are cavity-selected, that hyperbolicity brings no benefit, and that the nanolaminate-free stack is preferred.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Correlation‐Aware Uncertainty Quantification and Robustness Assessment of Dual One‐Dimensional Photonic‐Crystal Refractive‐Index Sensors: A Numerical StudyInternational Journal of Optics · 2026
- Surface and Interface Engineering in Integrated Photonic Sensors: Performance Trade-Offs, Stability, and BenchmarkingMicromachines · 2026
- Calibration Identifiability and Conditional Design Choices for a Dual-FBG Displacement SensorJournal of Sensor and Actuator Networks · 2026
- Reliability of Fisher-information bounds for dielectric metalenses: a site-resolved RCWA auditarXiv · 2026
- Tailoring Plasmonic Waveguide-Coupled Cavities for CO2 Sensing.ACS omega · 2026
- Theoretical investigation of an ultra-sensitive angular refractive index sensor based on compound parabolic micro-funnel arrays.Optics letters · 2026