Improving Wearable PPG Monitoring with Morphology-Aware and Learning-Driven Heartbeat Detection.
Pietro Chierico, Daniel Suárez, Lorenzo Fácila, Fernando Hornero, Raúl Alcaraz, José J Rieta
BioMIT.org, Electronic Engineering Department, Universitat Politècnica de València, 46022 Valencia, Spain. · Benejúzar Health Center, Orihuela Health Department, 03390 Benejúzar, Spain.
Key points
From the paper's conclusion- These results support PCPD as a promising PPG-only peak detector and motivate further validation under real motion artifacts and for downstream signal quality assessment.
Abstract
Photoplethysmography (PPG) is a widely used, non-invasive technique for heart rate monitoring, but its reliability is often compromised by noise and motion artifacts (MAs). This study presents the Peakwise Correlation Pulse Detector (PCPD), an algorithm for detecting heartbeats directly from PPG signals without requiring electrocardiogram (ECG) signals during operation, evaluated here under clean and controlled synthetic pulse-free-noise conditions. PCPD follows a five-step pipeline: (1) signal segmentation with overlap, (2) sliding cross-correlation using high-quality heartbeat templates, (3) computation of the Minimum Correlation Curve (MCC), (4) feature extraction with machine learning-based classification, and (5) peak translation with constraint analysis. The method was evaluated on a clean dataset, comprising a newly collected PPG dataset with manually annotated signal quality and the publicly available BIDMC dataset, and on a noisy dataset that augments the clean data with synthetic pulse-free segments generated from band-limited Gaussian noise. This evaluation framework tests heartbeat detectors under ideal conditions and under a controlled non-informative-noise protocol. On clean signals, PCPD achieved balanced precision-recall scores of 99.13% on the proposed dataset (internal performance) and 97.04% on the independent BIDMC dataset. Under added synthetic pulse-free noise, PCPD achieved accuracies of 99.30% and 97.16% on the proposed and BIDMC datasets, respectively. In comparative experiments against eleven benchmark heartbeat detection algorithms, PCPD achieved the best performance among the evaluated implementations under the tested synthetic-noise protocol. These results support PCPD as a promising PPG-only peak detector and motivate further validation under real motion artifacts and for downstream signal quality assessment.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A Novel Cluster-Based Approach for PPG Signal Preprocessing: Motion Artifact Filtering and Ambient Light Correction.Sensors (Basel, Switzerland) · 2026
- Reliable wrist PPG monitoring by mitigating poor skin sensor contactScientific Reports · 2025
- Robust PPG-Exclusive Heart Rate Estimation During Maximal Exercise Using Population-Based PriorsarXiv · 2026
- MorphSSF: a fast slope-sum adaptive-period dilation algorithm for photoplethysmography beat detection.Physiological measurement · 2026
- Continuous Cuffless Blood Pressure Estimation Using PPG From Methods to Clinical TranslationarXiv · 2026
- Photoplethysmography as a tool to assess heart rate variability and predict mortality in the general population: the INVADE studyEuropean Heart Journal · 2026