Fuzzy Fault Detection and Band-Conditioned Data Quality Assessment for Sensor Streams: A Continuous Glucose Monitoring Case Study.
Juan F Gaitán-Guerrero, Carmen Martínez-Cruz, Macarena Espinilla, José L López
Department of Computer Science, University of Jaén, Campus Las Lagunillas, 23071 Jaén, Spain. · Department of Languages and Computer Systems, University of Granada, 18071 Granada, Spain.
Key points
From the paper's conclusion- We propose a domain-agnostic, five-layer fuzzy pipeline that characterizes the reliability of a sensor stream jointly over time and value bands and carries it through to every derived aggregate: (L1) fault tagging of each sample against five fuzzy failure primitives; (L2) fuzzy windowing; (L3) band reliability; (L4) routed repair with propagated reconstruction uncertainty; and (L5) veracity aggreg
Abstract
Sensor faults can propagate beyond raw measurements and distort the aggregates used to interpret a monitored phenomenon. We propose a domain-agnostic, five-layer fuzzy pipeline that characterizes the reliability of a sensor stream jointly over time and value bands and carries it through to every derived aggregate: (L1) fault tagging of each sample against five fuzzy failure primitives; (L2) fuzzy windowing; (L3) band reliability; (L4) routed repair with propagated reconstruction uncertainty; and (L5) veracity aggregation with linguistic protoforms. An end-to-end acquisition architecture is also specified. The pipeline was instantiated for continuous glucose monitoring (CGM) with clinical knowledge of glucose dynamics and applied to a 14-month recording of an adult with type 1 diabetes (FreeStyle Libre 3, 5-min cadence), of which 294 days contained at least 120 readings. Clean reference days remained essentially unchanged (mean sample quality 0.98-0.99). On the most corrupted day, the estimated time below range fell from 16% to 3%, with a veracity of only 0.14. Over the recording, 38% of the days were too unreliable to phenotype and 5% changed phenotype. Joint time- and band-conditioned reliability assessment thus prevents corrupted sensor data from producing falsely precise clinical conclusions.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Distinguishing Process Change, Instrument Fault and Data-Path Corruption in Streaming Telemetry: A Reference Implementation and BenchmarkarXiv · 2026
- Benchmarking Hypoglycemia Classification Using Quality-Enhanced DiaDataIEEE Journal of Biomedical and Health Informatics · 2025
- Evaluation and Remediation of Data Quality for Drilling-Fluid Property Online Monitoring.Sensors (Basel, Switzerland) · 2026
- Uncertainty-aware fourier-encoded attentive Bi-GRU for non-invasive blood glucose estimation from multimodal PPG and GSR signals.Frontiers in medical technology · 2026
- Unsupervised Multivariate Anomaly Detection for Near-Real-Time Water-Quality Monitoring in District Metered Areas: A Three-Tier Benchmark Validated with Real-World AnomaliesWater · 2026
- Impact of Missing Data and Monitoring Duration on Downstream Analyses in Continuous Glucose MonitoringDiabetes Care · 2026