A Dual-Modal Hybrid Marker for Automatic Registration and Coordinate Hand-Over in C-Arm CT-Guided Spinal Navigation.
Taeho Kim, Sungmin Kim
Department of Electrical, Electronic, and Computer Engineering, University of Ulsan, Ulsan 44610, Republic of Korea. · School of Mechanical Engineering, University of Ulsan, Ulsan 44610, Republic of Korea.
Key points
From the paper's conclusion- These findings support the integrated registration and handover workflow as a basis for further validation at independent anatomical targets.
Abstract
An asymmetric hybrid registration frame (RF) integrates radiopaque and optical markers to automate computed tomography (CT) registration and coordinate hand-over to a dynamic reference frame (DRF). We compared automatic and manual fiducial selection using three fixed CT datasets of a rigid sawbone phantom. Ten operators performed five manual registrations per CT, alongside one automatic benchmark per CT. Both methods used the same five registration fiducials, rigid registration procedure, and three evaluation centers excluded from transform estimation. Target registration error (TRE) was the root mean square distance between transformed design centers and CT-detected centers, without probe contact. Across equally weighted CT datasets, mean TRE was 0.538 mm for automatic and 1.565 mm for manual registration. The manual-minus-automatic difference was 1.027 mm (95% operator-bootstrap confidence interval, 0.951-1.148 mm). Mean registration processing times were 2.817 and 22.589 s, respectively. Under these common evaluation conditions, automatic fiducial detection reduced on-frame center discrepancy and registration time. These findings support the integrated registration and handover workflow as a basis for further validation at independent anatomical targets.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Bridging the Markerless Tracking Validation Gap: Intraoperative X-ray Workflow IntegrationarXiv · 2026
- Visually detectable pointer deviation of the nextar surgical navigation system with c-arm, o-arm and ct imaging: a bench-top study using a two-dimensional phantom.BMC musculoskeletal disorders · 2026
- Comparative evaluation of CT-to-Fluoroscopy and Scan-and-Plan registration techniques in robot-assisted pedicle screw placement: a preliminary cadaveric faesibility study.Journal of robotic surgery · 2026
- The Best of Both Worlds: Combining Freehand and Mobile Intraoperative CT-Navigated Instrumentation in Pediatric Deformity Corrections—A Retrospective Case Series of 31 Consecutive PatientsSurgical Techniques Development · 2026
- Augmented reality-enabled image-guided surgical navigation in minimally invasive spine surgery: application beyond screw placement in prone lateral transpsoas interbody fusion-a cadaveric surgical technique study.Journal of spine surgery (Hong Kong) · 2026
- Accuracy of Augmented Reality-Guided Angular Trajectories in Spine Surgery: A Cadaveric Study Comparing 2D-3D vs Intraoperative Computed Tomography Registration.International journal of spine surgery · 2026