Design and Performance Validation of a Low-Loss Optical System with Light Source Fluctuation Compensation for In Situ Deep-Sea Nitrate Sensing.
Zhanfei Zhang, Jing Dong, Yahui Li, Yafeng Wu, Guojun Wu
Xi'an Institute of Optics and Precision Mechanics, Chinese Academy of Sciences, Xi'an 710119, China. · Laoshan National Laboratory, Qingdao 266237, China.
Key points
From the paper's conclusion- Performance evaluation of the system demonstrated a detection limit of 0.0056 mg/L-N(0.4 µmol/L, 10 mm optical path), with high measurement accuracy and precision across samples with varying nitrate concentrations.
Abstract
In situ measurement of deep-sea nitrate concentration is of significant importance for marine environmental monitoring and management. Nitrate sensors based on ultraviolet (UV) spectrophotometry, featuring simple, rapid, and reagent-free detection, are highly suitable for long-term autonomous monitoring and have attracted increasing research interest. However, existing systems still suffer from increased measurement errors and reduced detection ranges due to light source attenuation and optical transmission losses. To address these limitations, this study presents an optical system for a deep-sea nitrate sensor designed for deployment on mobile platforms. A co-aperture coupling design integrating the measurement and reference paths is proposed to achieve low optical loss while accurately compensating for the impact of light source attenuation on measurement accuracy. Furthermore, full-UV optical component optimization is performed to significantly improve optical energy utilization. Light-source calibration experiments demonstrate stable absorbance responses and effectively compensate for measurement deviations caused by light source attenuation. Performance evaluation of the system demonstrated a detection limit of 0.0056 mg/L-N(0.4 µmol/L, 10 mm optical path), with high measurement accuracy and precision across samples with varying nitrate concentrations. Sea trial results confirmed that the proposed system satisfies the requirements for stability and reliability in deep-sea in situ nitrate concentration monitoring, highlighting its potential for practical marine applications.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Polymer Optofluidic Chip for Double-Beam Photometric Determination of Nitrite in Water Samples.ACS omega · 2026
- Dual-Wavelength OA-ICOS for Sensitive and Wide-Range In Situ Measurement of Dissolved Methane in Deep-Sea Cold Seeps.Analytical chemistry · 2026
- Underwater Optical Imaging Systems: A Comprehensive Review of Technologies, Architectures and Mission-Driven Design Trade-Offs.Sensors (Basel, Switzerland) · 2026
- A Unified Channel Model and Performance Evaluation for Underwater-to-Air Optical LinksarXiv · 2026
- Sample-Specific Signal-to-Noise Ratio in Benchtop and Handheld NIR Reflectance SpectroscopyACS Measurement Science Au · 2026
- Photon-recycling lensless fluorometerJournal of the Optical Society of America B · 2026