Physics-Informed Design and Bench/Phantom Validation of a Shaft-Compatible 13.56 MHz NFC System for Laparoscopic Colorectal Tumour Localisation.
Bogdan Mocan, Mihaela Mocan, Mircea Fulea, Mircea Murar, Zsolt Mate, Adrian Calborean, Vasile V Bintintan
Department of Engineering Design and Robotics, Technical University of Cluj-Napoca, 400114 Cluj-Napoca, Romania. · Department of Internal Medicine, Iuliu Hațieganu University of Medicine and Pharmacy, 400012 Cluj-Napoca, Romania.
Key points
From the paper's conclusion- The study demonstrates a physics-informed route from antenna miniaturisation to measured system performance, and defines the present operating envelope under controlled bench and tissue-equivalent phantom conditions. The electromagnetic measurements apply to the antenna-electronics subassembly; integrated-shaft, multi-prototype, multi-operator, ex vivo, and in vivo validation remain necessary before clinical performance can be determined.
Abstract
BACKGROUND/OBJECTIVES: Accurate intraoperative tumour localisation remains challenging in minimally invasive colorectal surgery because tactile palpation is lost and conventional markers can migrate or provide imprecise localisation. Building on a preceding tri-frequency study that identified 13.56 MHz as the preferred RFID band for the intended application, this work develops a shaft-compatible NFC antenna-reader platform and evaluates its electromagnetic behaviour from bench-top reference media to five-layer tissue-equivalent phantoms. METHODS: A Ø3 × 25 mm Fair-Rite Material 67 ferrite-rod antenna was designed from material and geometric parameters using finite-rod demagnetisation, inductance, resonance, and field calculations, followed by FEM cross-validation and experimental characterisation. The primary dataset comprised 480 detection distance measurements (2 media × 4 tag angles × 30 repetitions × 2 encapsulation variants). Phantom testing added 1440 measurements at 22 °C and 600 measurements at 37 °C across three fabrication batches, with the 37 °C non-coaxial subset limited to one batch. RESULTS: The fabricated antenna measured 16.9 µH versus a 17.4 µH analytical estimate (-2.9%), with loaded Q = 23. The coaxial detection range was 16.45 ± 0.29 mm in air and 16.26 ± 0.21 mm in saline; angle was the dominant determinant of range (partial η2 = 0.989). In the multi-layer phantom, detection was 100% at 0 and 10 mm perirectal fat thickness under coaxial alignment at 22 °C, whereas performance declined markedly with angular misalignment and no detections occurred at fat thicknesses ≥ 20 mm. Across detectable phantom configurations, FEM showed r2 = 0.994, RMSE = 0.81 mm, and mean bias +0.70 mm. Bare and resin-overcoated tags showed no statistically detectable range difference. Multi-tag discrimination reached 100% for up to three tags separated by ≥20 mm under coaxial alignment, but deteriorated with angular misalignment. CONCLUSIONS: The study demonstrates a physics-informed route from antenna miniaturisation to measured system performance, and defines the present operating envelope under controlled bench and tissue-equivalent phantom conditions. The electromagnetic measurements apply to the antenna-electronics subassembly; integrated-shaft, multi-prototype, multi-operator, ex vivo, and in vivo validation remain necessary before clinical performance can be determined.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A Compact Non-Ionizing RF Bioelectronic Sensor for Translational Phantom-Based Detection of Tumor-Like Dielectric Changes in Bone.Nanotheranostics · 2026
- An FBG-Based Miniature Force-Sensing Retractor for In Vivo Measurement of Rat Lumbar Nerve Root Traction Force and Injury Assessment.Micromachines · 2026
- Comparison of different phantoms for teaching ultrasound‐guided fascia iliaca block—Users surveyHong Kong Journal of Emergency Medicine · 2026
- Endoscopic NIRAF-LSCI multimodal system with pseudo-color visualization for parathyroid autofluorescence localization and tissue perfusion assessment: preclinical validation.Lasers in medical science · 2026
- Robust inverse dielectric design of skin phantoms for implantable wireless link evaluationarXiv · 2026
- A Compact Ultra‐Wideband Implantable Antenna for Bovine Health MonitoringMicrowave and Optical Technology Letters · 2026