An FBG-Based Miniature Force-Sensing Retractor for In Vivo Measurement of Rat Lumbar Nerve Root Traction Force and Injury Assessment.
Jixin Li, Jun Miao, Jie Lan, Min Wang, Bangquan Liao
Department of Orthopedics, Beichen Hospital Affiliated to Nankai University, Tianjin 300400, China. · Department of Spine Surgery, Tianjin Hospital, Tianjin University, Tianjin 300211, China.
Key points
From the paper's conclusion- This study validates the feasibility of an FBG-based miniature force-sensing retractor for quantitative in vivo measurement of nerve root traction force and provides an experimental tool for investigating the biomechanical relationship between traction loading and nerve root injury responses.
Abstract
Miniature tools for quantitative measurement of nerve root traction force during posterior lumbar surgery remain limited, restricting objective evaluation of the relationship between traction loading and nerve root injury. In this study, we developed a miniature force-sensing retractor based on fiber Bragg grating (FBG) technology for in vivo measurement of rat lumbar nerve root traction force. The FBG sensing element was encapsulated on the loading side of the miniature retractor and calibrated using standard weight loading. A displacement-controlled rat lumbar nerve root traction model was established to simulate lateral nerve root retraction during posterior lumbar surgery. Rats were assigned to Sham, Mild, Moderate, and Severe groups with predefined lateral displacements of 0, 0.36, 0.72, and 1.08 mm, respectively, maintained for 10 min. Mechanical calibration demonstrated a linear FBG response using an origin-constrained regression model, with a sensitivity of 3.62 nm/N and a theoretical force resolution of approximately 0.006 N. The group-representative steady-state traction forces corresponding to the Mild, Moderate, and Severe groups were approximately 0.09, 0.22, and 0.45 N, respectively. Different traction levels were associated with functional and histological changes at postoperative Day 28. The Severe group showed persistent myelin damage, increased F4/80-positive area, and reduced NF-200-positive area. This study validates the feasibility of an FBG-based miniature force-sensing retractor for quantitative in vivo measurement of nerve root traction force and provides an experimental tool for investigating the biomechanical relationship between traction loading and nerve root injury responses.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Fiber Bragg grating-based three-dimensional human-robot interaction force sensor with high precision and large range for lower limb rehabilitation exoskeleton.Optics express · 2026
- PALPABLE: Real‐Time Soft Tissue Stiffness Detection and Mapping Using Multiplexed Fiber Bragg GratingsAdvanced Robotics Research · 2026
- Vibration-Based Measurement of Steel Wire Rope Fundamental Frequency Using an Optical Sensor: Mitigation of Mass-Loading Artifacts.Sensors (Basel, Switzerland) · 2026
- A Fiber Bragg Grating-Based Measurement Method for Outer-Ring Fault Detection in Rolling Bearings.Sensors (Basel, Switzerland) · 2026
- Motion Compensation and Adaptive Force Control via iOCT-FBG Sensor Fusion for Robotic Subretinal Injection.IEEE International Conference on Robotics and Automation : ICRA : [proceedings]. IEEE International Conference on Robotics and Automation · 2026
- A Brace-Integrated Fiber Bragg Grating Sensor with a Multi-Hardness Silicone Strain Transfer Substrate for Knee Flexion Monitoring: A Proof-of-Concept Study.Sensors (Basel, Switzerland) · 2026