Sensing in Urinary Catheter Systems: From Measurands and Signal Fidelity to Supportable Clinical Claims.
Shuai Zhang
School of Pharmacy, Queen's University Belfast, Belfast BT9 7BL, UK.
Key points
From the paper's conclusion- Translation requires validation over complete catheter episodes against an endpoint matched to the intended claim.
Abstract
This review evaluates urinary catheter sensors designed to warn of crystalline encrustation and blockage or to monitor infection-related microbial, interfacial, and host-response states. A source-to-claim framework links each measured state and sampling compartment with its readout, reference endpoint, and supportable clinical claim. The analysis also considers whether catheter integration preserves the relationship between source, readout, and claim. pH-responsive systems have the strongest human feasibility evidence, although the studies are small, use different sampling configurations, and include few blockage events. Evidence for urease sensing is confined to analytical studies and controlled in vitro experiments. Mineral-deposition and hydraulic approaches rely mainly on catheter models or adjacent-device studies. Most infection-related platforms are sample-based assays, prototypes, or short-duration systems that measure clinically non-equivalent states. Analyte detection alone cannot establish a diagnosis or predict a later event in either application. Interpretation also changes with sensor location and with the effects of transport, fouling, and mechanical loading on the recognition interface. Translation requires validation over complete catheter episodes against an endpoint matched to the intended claim. An additional channel is warranted only if it resolves a prespecified uncertainty and improves clinically relevant performance against a fixed comparator.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Safety, usability, patient comfort, and exploratory biofilm outcomes of the superhydrophilic antifouling LubriShield(R) Foley catheter: a randomised controlled trialmedRxiv · 2026
- Remote Coacervate Robots with Visualized Bacteria Sensing and Removal in Indwelling Catheters for Infection Prevention.Nano letters · 2026
- Empirical antibiotic use in nonsurgical patients with indwelling urinary catheters: a systematic reviewFuture Microbiology · 2026
- Clinical Effectiveness of Tap Water Versus Povidone-Iodine Bladder Irrigation for Recurrent Catheter-Associated Urinary Tract Infections: A Comparative Review of Contemporary Clinical Evidence.Cureus · 2026
- Zwitterionic Coatings for Urinary Catheters: Current Progress, Design Principles and Translational Challenges.ACS omega · 2026
- Wireless MEMS pressure sensing for intrarenal pressure monitoring during ureteroscopy: an in vitro bench-feasibility study.Frontiers in bioengineering and biotechnology · 2026